Evaluation of the dosimetric response for CaSO4:Dy at low temperature.

Radiat Prot Dosimetry

Instituto de Ciencias Nucleares, UNAM, México, DF México.

Published: April 2003

Homemade solid state CaSO4:Dy detectors were tested to evaluate their response to gamma radiation at liquid nitrogen temperature (77 K). The dosemeters were irradiated with doses between 12 and 1071 Gy. For this study these dosemeters were exposed to gamma rays with a dose rate of 1.19 Gy.min(-1). The analysis for these crystals was made by thermoluminiscence. The dose response at liquid nitrogen temperature was linear in the dose range studied and it is about 20% lower with respect to the response at room temperature. The response is reproducible with the same geometric set-up.