Purpose: Determination of image quality between conventional film screen system, digital phosphor storage plate mammography (high resolution) and digital mammography.
Materials And Methods: Mammograms of the Wisconsin Mammographic Random Phantom, Model 152 A (Radiation Measurements Inc., Wisconsin) were acquired using a conventional film-screen system, a digital storage phosphor plate system and a digital system.
Results: Of 225 possible details, 191/a 38.2, 193/a 38.6, and 202/a 40.4 details were detected with conventional film-screen system, digital phosphor storage plate mammography and digital mammography, respectively. There was no significant difference (p < 0,058). The entrance surface air kerma was 9.64 mGy, 7.60 mGy and 7.02 mGy, respectively.
Conclusions: Based on these results, conventional film-screen system can be replaced with both digital phosphor storage plate mammography and digital mammography, to be confirmed with further clinical trials.