Unlabelled: The NU 2-1994 standard document for PET performance measurements has recently been updated. The updated document, NU 2-2001, includes revised measurements for spatial resolution, intrinsic scatter fraction, sensitivity, counting rate performance, and accuracy of count loss and randoms corrections. The revised measurements are designed to allow testing of dedicated PET systems in both 2-dimensional and 3-dimensional modes as well as coincidence gamma cameras, conditions not considered in the original NU 2-1994 standard. In addition, the updated measurements strive toward being more representative of clinical studies, in particular, whole-body imaging.
Methods: Performance measurements following the NU 2-1994 and NU 2-2001 standards were performed on several different PET scanners. Differences between the procedures and resulting performance characteristics, as well as the rationale for these changes, were noted.
Results: Spatial resolution is measured with a point source in all 3 directions, rather than a line source, as specified previously. For the measurements of intrinsic scatter fraction, sensitivity, and counting rate performance, a 70-cm line source is now specified, instead of a 19-cm-long cylindric phantom. The longer configuration permits measurement of these performance characteristics over the entire axial field of view of all current PET scanners and incorporates the effects of activity outside the scanner. A measurement of image quality has been added in an effort to measure overall image quality under clinically realistic conditions. This measurement replaces the individual measurements of uniformity and of the accuracy of corrections for attenuation and scatter.
Conclusion: The changes from the NU 2-1994 standard to the NU 2-2001 standard strive toward establishing relevance with clinical studies. The tests in the updated standard also are, in general, simpler and less time-consuming to perform than those in the NU 2-1994 standard.
Download full-text PDF |
Source |
---|
Med Phys
January 2024
Department of Radiology, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
Background: Scatter correction (SC) is essential in PET for accurate quantitative imaging. The state-of-the-art SC method is single-scatter simulation (SSS). Although this method is usually robust and accurate, it can fail in some situations, for example when there is motion between the CT and PET scans in PET/CT.
View Article and Find Full Text PDFMaterials (Basel)
June 2023
Research and Development Center, Atlas sp. z o.o., 2 Kilińskiego St., 91-421 Lodz, Poland.
The paper presents the results of tensile adhesion strength measurements of ceramic tile adhesive (CTA) stored in various conditions performed by ten operators in one laboratory using the same equipment and auxiliary materials. The obtained results allowed the authors to estimate the repeatability and reproducibility of the tensile adhesion strength measurement method using the methodology following ISO 5725-2:1994+AC1:2002. Standard deviations of repeatability ranging from 0.
View Article and Find Full Text PDFJ Microbiol Methods
May 2021
French Agency for Food, Environmental and Occupational Health and Safety (ANSES), European Union Reference Laboratory for Bee Health, Honey Bee Pathology Unit, 105 route des Chappes - CS 20111, 06902 Sophia Antipolis, France.
Nosemosis is a microsporidian disease causing mortality and weakening of honey bee colonies, especially in the event of co-exposure to other sources of stress. As a result, the disease is regulated in some countries. Reliable and harmonised diagnosis is crucial to ensure the quality of surveillance and research results.
View Article and Find Full Text PDFFood Chem X
June 2020
European Commission, Joint Research Centre (JRC), Retieseweg 111, B 2440 Geel, Belgium.
Meat from male pigs may develop an off-flavour, commonly known as boar taint. Castration of male piglets prevents the potential formation of off-flavour. In the suggested method, three marker compounds for boar taint (skatole, androstenone and indole) are quantified in pork fat by isotope dilution gas chromatography mass spectrometry (GC-MS) or by isotope dilution liquid chromatography tandem mass spectrometry (LC-MS/MS).
View Article and Find Full Text PDFJ Virol Methods
October 2017
French Agency for Food, Environmental and Occupational Health & Safety (ANSES), European Reference Laboratory for Honeybee Health, Unité pathologie de l'abeille, CS20111, 06902 Sophia Antipolis, France. Electronic address:
The Chronic bee paralysis virus (CBPV) is the aetiological agent of chronic bee paralysis, a contagious disease associated with nervous disorders in adult honeybees leading to massive mortalities in front of the hives. Some of the clinical signs frequently reported, such as trembling, may be confused with intoxication syndromes. Therefore, laboratory diagnosis using real-time PCR to quantify CBPV loads is used to confirm disease.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!