Intra-aortic balloon (IAB) counterpulsation can augment the cardiac output. However, the effect of different IAB volumes on cardiac performance has not been adequately evaluated in humans. Eighty-two patients (52 males [63%]; mean age, 65 +/- 12 years; mean body surface area [BSA], 1.8 +/- 0.2 m(2)) had IAB counterpulsation for cardiogenic shock, refractory angina, and preoperatively for high-risk cardiac surgery. Cardiac hemodynamics were prospectively studied during IAB with inflation volumes of 32 vs. 40 cc. Hemodynamic data collected included aortic pressure, pulmonary artery pressure, systemic and mixed venous oxygen saturations, and cardiac output (by Fick). Transthoracic echocardiography (TTE) was used to obtain both velocity time integrals (VTIs) and the area of the left ventricular outflow tract (LVOT). Left ventricular stroke volume was then calculated as LVOT area x VTI. Cardiac output (CO) determined by the Fick method and VTI did not differ significantly (P = NS) between the two inflation volumes (y = 0.002 + 0.97x). In a subgroup of 33 patients with BSA

Download full-text PDF

Source
http://dx.doi.org/10.1002/ccd.10304DOI Listing

Publication Analysis

Top Keywords

cardiac output
12
intra-aortic balloon
8
volumes cardiac
8
cardiac performance
8
iab counterpulsation
8
inflation volumes
8
left ventricular
8
cardiac
7
impact intra-aortic
4
balloon counterpulsation
4

Similar Publications