Technical reliability of the CSA activity monitor: The EarlyBird Study.

Med Sci Sports Exerc

University Medicine Level 7, Derriford Hospital, University of Plymouth, Plymouth PL6 8DH, UK.

Published: September 2002

Purpose: To evaluate the technical performance of the CSA accelerometer-based activity monitor.

Methods: Twenty-three CSA monitors were subjected to intra- and inter-instrument variability tests by controlled trials using a motorized turntable. The CSA monitor measures change in acceleration, and precision was tested by producing sinusoidal variations in speed around two fixed baseline speeds (fast and medium). The angle of the monitor to the line of force along the radius of the turntable was varied using tilted blocks. Three sets of tests were carried out. 1. Intra-instrument variability: seven monitors were tested three times in each of the four quadrants. 2. All 23 monitors were used for inter-instrument tests. 3. The effects of tilt at 15 degrees, 30 degrees, and 45 degrees were carried out on six monitors.

Results: Intra-instrument coefficients of variation (CV) never exceeded 2% for fast or medium speed and achieved "between run" intra-class correlation coefficients (ICC) of 0.92 and 0.84 respectively. There were no significant differences between the monitors in terms of repeatability (fast: = 0.97, medium: = 0.77). Although there were significant differences between monitors in terms of mean score, inter-instrument variability did not exceed 5% at either speed. Inter-batch ICCs ranged from 0.87 to 0.98 for fast and from 0.71 to 0.99 for medium. The angle test results corresponded closely to those predicted theoretically, with a loss in mean score of only 6% when the monitor was tilted from 0 degrees to 15 degrees.

Conclusion: The CSA monitor provides a precise tool for measuring changes in acceleration in laboratory settings. Technically, the device performs well, and is likely to prove a useful tool in the assessment of physical activity in children and adults.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00005768-200209000-00022DOI Listing

Publication Analysis

Top Keywords

inter-instrument variability
8
csa monitor
8
fast medium
8
medium angle
8
degrees degrees
8
differences monitors
8
monitors terms
8
csa
5
monitor
5
monitors
5

Similar Publications

Efforts to simplify standard polysomnography (PSG) in laboratories, especially for obstructive sleep apnea (OSA), and assess its agreement with portable electroencephalogram (EEG) devices are limited. We aimed to evaluate the agreement between a portable EEG device and type I PSG in patients with OSA and examine the EEG-based arousal index's ability to estimate apnea severity. We enrolled 77 Japanese patients with OSA who underwent simultaneous type I PSG and portable EEG monitoring.

View Article and Find Full Text PDF

Detection and agreement of event-based OCT and OCTA analysis for glaucoma progression.

Eye (Lond)

April 2024

Hamilton Glaucoma Center, Shiley Eye Institute, Viterbi Family Department of Ophthalmology, University of California, San Diego, La Jolla, CA, USA.

Objective: To examine event-based glaucoma progression using optical coherence tomography (OCT) and OCT angiography (OCTA).

Methods: In this retrospective study, glaucoma eyes with ≥2-year and 4-visits of OCT/OCTA imaging were included. Peripapillary capillary density (CD) and retinal nerve fibre layer thickness (RNFL) were obtained from 4.

View Article and Find Full Text PDF

The Peppermint Experiment is a breath analysis benchmarking initiative that seeks to address the lack of inter-comparability of outcomes across independent breath biomarker studies. In this experiment, the washout profiles of volatile terpene constituents of encapsulated peppermint oil (mainly-pinene,-pinene, limonene and 1,8-cineole) in exhaled breath are characterized through a series of measurements at defined sampling intervals up to 6 h after ingestion of the capsule. In the present work, the Peppermint Experiment was carried out on a cohort of volunteers (= 11) that provided breath samples in three sittings on different days (i.

View Article and Find Full Text PDF

The measurement of fine (diameter: 100 nanometers-2.5 micrometers) and ultrafine (UF: < 100 nanometers) titanium dioxide (TiO) particles is instrument dependent. Differences in measurements exist between toxicological and field investigations for the same exposure metric such as mass, number, or surface area because of variations in instruments used, operating parameters, or particle-size measurement ranges.

View Article and Find Full Text PDF

Inter-intra instrument comparison and standardization of a 10-color immunophenotyping for B and T cell non-Hodgkin lymphoma diagnosis and monitoring.

J Immunol Methods

December 2022

Hematology and Transplant Center, University Hospital "San Giovanni di Dio e Ruggi d'Aragona", Salerno, Italy; Department of Medicine, Surgery, and Dentistry, University of Salerno, Baronissi, Italy.

Harmonization of flow cytometry protocols from instrument settings to antibody panel and reagents is highly encouraged for inter-laboratory data comparison in both research and clinical settings, especially for minimal residual disease monitoring evaluation in hematological diseases across centers. Here, we described inter-intra instrument comparison of two standardized 10-color staining dried tubes for B- and T-cell lymphoproliferative disorder diagnosis and monitoring on two different flow cytometers, a Beckman Coulter NaviosEx and a Beckman Coulter DxFlex. A total of 47 consecutive patients were enrolled, and 39 of them were evaluable for further studies.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!