Objective: To determine the relative stability of three fixation methods for displaced capitellum fractures.

Design: Twelve matched pairs of embalmed humeri were divided into two equal groups and simulated capitellum fractures created. The first group compared cancellous lag screws placed in an anteroposterior direction to screws placed in the posteroanterior direction. The second group compared the Acutrac compression screw, inserted anteroposteriorly, to the more stable construct from the first test group.

Methods: All specimens were cyclically tested with simulated physiologic loading. Both displacement of the capitellum over a range of cycles and the number of cycles to failure were recorded.

Results: Fixation with posteroanteriorly directed cancellous lag screws was significantly more stable than anteroposteriorly directed screws at 2000 cycles (p = 0.007); loads to failure were not statistically different. Fixation by the Acutrac screws was significantly more stable than posteroanterior cancellous screws at 2000 cycles (p = 0.03). The Acutrac fixation had a higher failure load; however, this was not statistically significant.

Conclusion: The headless screws tested in this biomechanical study provided more stable fixation of capitellum fractures in the cadaveric specimens than four-millimeter partially threaded cancellous lag screws and may do so in the clinical setting. When the cancellous lag screws were tested, insertion in the posteroanterior direction provided more stable fixation than the anteroposterior direction and has clinical benefit of not violating the articular surface. Ultimately, the decision of which method to use lies with the attending surgeon and the technique with which he or she feels most comfortable.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00005131-200208000-00009DOI Listing

Publication Analysis

Top Keywords

cancellous lag
16
lag screws
16
capitellum fractures
12
screws
9
three fixation
8
fixation methods
8
group compared
8
anteroposterior direction
8
posteroanterior direction
8
screws stable
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!