Background: Uptake of a second gas of a delivered gas mixture decreases the amount of carrier gas and potent inhaled anesthetic leaving the circle system through the pop-off valve. The authors hypothesized that the vaporizer settings required to maintain constant end-expired sevoflurane concentration (Etsevo) during minimal-flow anesthesia (MFA, fresh gas flow of 0.5 l/min) or low-flow anesthesia (LFA, fresh gas flow of 1 l/min) would be lower when sevoflurane is used in oxygen-nitrous oxide than in oxygen.

Methods: Fifty-six patients receiving general anesthesia were randomly assigned to one of four groups (n = 14 each), depending on the carrier gas and fresh gas flow used: group Ox.5 l (oxygen, MFA), group NOx.5 l (oxygen-nitrous oxide, MFA after 10 min high fresh gas flow), group Ox1 l (oxygen, LFA), and group NOx1 l (oxygen-nitrous oxide, LFA after 10 min high fresh gas flow). The vaporizer dial settings required to maintain Etsevo at 1.3% were compared between groups.

Results: Vaporizer settings were higher in group Ox.5 l than in groups NOx.5 l, Ox1 l, and NOx1 l; vaporizer settings were higher in group NOx.5 l than in group NOx1 l between 23 and 47 min, and vaporizer settings did not differ between groups Ox1 l and NOx1 l.

Conclusions: When using oxygen-nitrous oxide as the carrier gas, less gas and vapor are wasted through the pop-off valve than when 100% oxygen is used. During MFA with an oxygen-nitrous oxide mixture, when almost all of the delivered oxygen and nitrous oxide is taken up by the patient, the vaporizer dial setting required to maintain a constant Etsevo is lower than when 100% oxygen is used. With higher fresh gas flows (LFA), this effect of nitrous oxide becomes insignificant, presumably because the proportion of excess gas leaving the pop-off valve relative to the amount taken up by the patient increases. However, other unexplored factors affecting gas kinetics in a circle system may contribute to our observations.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00000542-200208000-00017DOI Listing

Publication Analysis

Top Keywords

fresh gas
24
vaporizer settings
20
gas flow
20
oxygen-nitrous oxide
20
gas
14
carrier gas
12
pop-off valve
12
required maintain
12
low-flow anesthesia
8
circle system
8

Similar Publications

This study investigated the effects of different protein sources on feed intake, nutrient, and energy utilization, growth performance, and enteric methane (CH4) emissions in growing beef cattle, also evaluated against a pasture-based diet. Thirty-two Holstein × Angus growing beef were allocated to four dietary treatments: a total mixed ration (TMR) including solvent-extracted soybean meal as the main protein source (SB; n = 8), TMR with local brewers' spent grains (BSG; n = 8), TMR with local field beans (BNS; n = 8), and a diet consisting solely of fresh-cut Italian ryegrass (GRA; n = 8). Every four weeks, animals were moved to digestibility stalls within respiration chambers to measure nutrient intakes, energy and nitrogen (N) utilization, and enteric CH4 emissions.

View Article and Find Full Text PDF

Anesthetic gases contribute to global warming. We described a two-year performance improvement project to examine the association of individualized provider dashboard feedback of anesthetic gas carbon dioxide equivalent (CDE) production and median perioperative fresh gas flows (FGF) during general anesthetics during perioperative management. Using a custom structured query language (SQL) query, hourly CDE for each anesthetic gas and median FGF were determined.

View Article and Find Full Text PDF

Necrotizing enterocolitis (NEC) is a devastating disease of the neonatal gastrointestinal tract. Volatile organic compounds (VOCs), odoriferous compounds released as a byproduct of bacterial metabolism, can be used as a proxy for gut health. We hypothesized that patients with NEC would have different microbial profiles and elicit different VOC signatures as assessed by gas chromatography/mass spectrometry (GC/MS) or an electronic nose compared to controls.

View Article and Find Full Text PDF

Wildfires at the wildland-urban interface (WUI) have been increasing in frequency over recent decades due to increased human development and shifting climatic patterns. The work presented here focuses on the impacts of a WUI fire on indoor air using field measurements of volatile organic compounds (VOCs) by Proton-Transfer-Reaction Time-of-Flight Mass Spectrometry (PTR-TOF-MS). We found a slow decrease in VOC mixing ratios over the course of roughly 5 weeks starting 10 days after the fire, and those levels decreased to ∼20% of the initial indoor value on average.

View Article and Find Full Text PDF

, a special economic aquaculture species in China, is valued highly for its medicinal and nutritional benefits. However, the muscle of farmed exhibits a strong off-flavor, resulting in poor flavor quality. To enhance the flavor quality of the meat, this study examined the volatile compounds in muscle by establishing identification methods for these volatile odor compounds and comparing the differences between the two aquaculture modes.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!