Study Objectives: To investigate the frequency of microarousals (MA) associated with pressure changes during auto-CPAP therapy (APAP) for obstructive sleep apnea (OSA).
Design: Patients with OSA were studied by polysomnography during APAP therapy (Somnosmart). The MA were classified on the basis of concomitant changes in APAP pressure.
Setting: Sleep laboratory of a university hospital
Participants: 30 patients with moderate to severe OSA.
Measurements And Results: The mean AHI during APAP was 4.7+/-4.7, the mean arousal index was 14.5+/-6.6 per hour. During epochs with a pressure variation greater than 0.5 mbar, significantly more MA occurred (0.30+/-0.17 MA per epoch) than in epochs with constant treatment pressure (0.10+/-0.054 MA per epoch; p<0.001). There were more MA during pressure-increase epochs than during pressure-decrease epochs (0.42+/-0.24 vs. 0.16+/-0.12 MA per epoch; p<0.001). 82.5 percent of the MA were not preceded by a significant change in pressure (at least 0.5 mbar within 30 sec.), 10.6% were associated with a significant prior increase and 6.9% with a significant prior decrease in pressure. The percentage of MA preceded by a significant pressure variation varied between 2.3% and 61%, with a mean of 18.9%.
Conclusions: The overall frequency of MA was low, and in most individuals the relative amount of "pressure-associated MA" was not significant. However in some individuals it cannot be excluded that some additional MA may have been induced by pressure variations. Should it prove possible to prevent such "pressure-associated MA" by optimizing the regulation of APAP pressure, the overall clinical effect of APAP treatment may be improved.
Download full-text PDF |
Source |
---|
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!