Purpose: To compare a magnetization-prepared gradient-echo (GRE) sequence with a conventional GRE sequence for visualizing contrast agent-filled catheters.

Materials And Methods: Passive visualization of endovascular catheters using MRI was compared between two imaging sequences: 1) inversion recovery (IR)-fast low angle shot (FLASH), and 2) conventional FLASH. Two-dimensional projection images of the catheters filled with 4% diluted contrast agent in a phantom and the aorta of swine were obtained with each sequence with a temporal resolution of two frames per second. We compared background suppression and catheter visibility using the catheter-to-background signal ratio and the ratings of two radiologists.

Results: In the phantom, IR-FLASH allowed for a 200% increase in catheter-to-background ratio (p < 0.01) and improved depiction of catheters over conventional FLASH. In swine, the IR-FLASH images showed a statistically significant improvement of 80% (p < 0.001) over conventional FLASH in all comparisons of the catheter-to-background signal ratio, and an improvement of 160% (p < 0.05) in comparison with the radiologists' observations.

Conclusion: This study shows that IR-FLASH is a better technique for passive tracking of contrast agent-filled catheters than conventional FLASH.

Download full-text PDF

Source
http://dx.doi.org/10.1002/jmri.10132DOI Listing

Publication Analysis

Top Keywords

conventional flash
16
gre sequence
8
contrast agent-filled
8
catheter-to-background signal
8
signal ratio
8
catheters conventional
8
conventional
6
flash
6
passive catheter
4
catheter tracking
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!