We report a case of a patient with symptomatic pacemaker-mediated tachycardia (PMT). PMT should be suspected when asynchronous pacing induced by magnet application causes temporary resolution of an unknown tachycardia in a patient with a pacemaker. The underlying mechanism of PMT is either reentrant tachycardia or atrial triggering. PMT is better discerned with an intracardiac electrocardiogram (ECG) obtained through pacemaker interrogation (PI) than the standard 12-lead ECG. In the case reported, the interrogator was used to demonstrate that the initiating dysrhythmia was atrial flutter. Next, the programmer was used to overdrive the atrium restoring normal sinus rhythm. After pacemaker reprogramming, the patient was discharged from the emergency department (ED) in improved condition. We believe this to be the first reported case of a patient with PMT diagnosed, treated, and discharged effectively after PI, overdrive pacing and reprogramming from the ED. The pacemaker interrogator/programmer is a rapid noninvasive device that has practical applications in the ED.

Download full-text PDF

Source
http://dx.doi.org/10.1053/ajem.2002.33780DOI Listing

Publication Analysis

Top Keywords

pacemaker-mediated tachycardia
8
case patient
8
pacemaker
5
pmt
5
tachycardia management
4
management pacemaker
4
pacemaker interrogation/reprogramming
4
interrogation/reprogramming report
4
report case
4
patient
4

Similar Publications

Article Synopsis
  • The study explores a new type of heart rhythm issue called pseudo-RNRVAS that occurs when atrial stimulation stops while there is continued communication between the ventricles and atrium during certain heart phases.
  • Researchers investigated 21 patients with cardiac devices, finding that 20 experienced pseudo-RNRVAS, particularly at lower heart rates and specific timing intervals after losing atrial capture.
  • The findings suggest that pseudo-RNRVAS can happen without the usual risk factors and are more likely in scenarios that increase atrial pacing, especially soon after pacemaker implantation, indicating a possible link to pacemaker syndrome and heart failure risk.
View Article and Find Full Text PDF
Article Synopsis
  • A 55-year-old man with nonischemic cardiomyopathy experienced a unique tachycardia recorded by an implantable cardioverter defibrillator, raising concerns about its identification and management.
  • The report outlines how to differentiate this type of tachycardia from similar heart rhythms and discusses the clinical implications related to patient care.
  • It also highlights the limitations of the device in handling these tachycardias and suggests reprogramming strategies to reduce the chances of recurrence, stressing the need for thorough analysis of device-related arrhythmias.
View Article and Find Full Text PDF
Article Synopsis
  • Endless loop tachycardia (ELT) is a frequent issue in patients with implanted cardiac devices and can cause serious symptoms despite preventive measures.
  • * A study analyzed 20 cases of sustained ELT from 4,520 device checkups over four years, finding it usually benign but occasionally harmful, particularly in patients with left ventricular dysfunction.
  • * Various factors contributed to ELT, including improper AV synchrony and sensing issues, and adjustments like changing sensitivity settings or PVARP duration were necessary for management.
View Article and Find Full Text PDF
Article Synopsis
  • Pacemaker-mediated tachycardia (PMRA) can occur in patients with dual chamber pacemakers and is particularly noted in those with intact ventriculoatrial conduction.
  • A 91-year-old female with new cardiomyopathy received an AV synchronous leadless pacemaker, which led to frequent atrial and ventricular pacing issues, identified as PMRA due to changes in heart rates.
  • The study highlights that PMRA may be overlooked in patients with leadless pacemakers, emphasizing the need for further research on its management and impact.
View Article and Find Full Text PDF

What are those intermittent "S" waves?

J Electrocardiol

June 2024

Department Internal Medicine, Atrium Health Carolinas Medical Center, Charlotte, NC, USA. Electronic address:

Article Synopsis
  • A 69-year-old woman experienced three fainting spells while on an airplane and was diagnosed with severe bradycardia, indicated by her low heart rate in the thirties.!
  • Two ECGs were performed: the first showed narrow QRS complexes, while the second displayed wide negative deflections after the QRS complexes.!
  • Telemetry recordings allowed doctors to identify the problem's root cause and properly program her implanted pacemaker for treatment.!
View Article and Find Full Text PDF