We compared an assay using signal amplification of a heat-dissociated p24 antigen (HDAg) with the Roche Monitor human immunodeficiency virus (HIV) RNA assay. The two assays gave comparable results when 130 specimens from 130 patients were tested (r = 0.60, P < 0.0001). The HDAg assay was almost as sensitive (85%) as the Roche HIV RNA kit (95%), just as specific (25 negative results from 25 HIV seronegative volunteers [100%]), less variable (mean log standard deviation of 0.07 compared to 0.11 when eight specimens were tested three or four times), and less expensive (reagent and labor costs, $8 versus $75). The assay appeared to be useful for monitoring established patients (n = 17) and identifying seroconverters (n = 4). HIV subtypes A to F were all recognized. This assay should be useful for monitoring patients in resource-poor countries and for monitoring vaccine recipients.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC120606 | PMC |
http://dx.doi.org/10.1128/JCM.40.7.2472-2475.2002 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!