[Neon--a new angle-stable implant system for dorsal occipitocervical instrumentation. Biomechanical comparison with established systems].

Orthopade

Rehabilitationskrankenhaus, Orthopädische Klinik, Querschnittgelähmtenzentrum, Universität Ulm, Oberer Eselsberg 45, 89081 Ulm.

Published: April 2002

Posterior instrumentation of the occipitocervical spine is well-established for different indications. The aim of this study was to evaluate whether posterior internal fixation of the occipitocervical spine with the new implant system improves primary biomechanical stability. Primary stability was significantly increased in all load cases with the new modular implant system compared to the other implant systems. Pedicle screw instrumentation tended to be stabler compared to lateral mass screws; nevertheless, significant differences could be observed only for lateral bending. As the experimental design precluded any cyclic testing, the data represent only the primary stability of the implants. In summary, this study showed that posterior instrumentation of the cervical spine using the new neon occipito-cervical system improves primary biomechanical stability compared to the CerviFix system and the Olerud cervical rod spinal system.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00132-001-0274-9DOI Listing

Publication Analysis

Top Keywords

implant system
12
posterior instrumentation
8
occipitocervical spine
8
system improves
8
improves primary
8
primary biomechanical
8
biomechanical stability
8
primary stability
8
system
6
[neon--a angle-stable
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!