An online automatic sample cleanup system was developed for use with electrospray ionization tandem mass spectrometry (ESI-MS-MS) for the quantitative detection of the benzene exposure biomarker S-phenylmercapturic acid (S-PMA) in human urine. The sample clean-up system was constructed with an autosampling device, a reversed-phase C18 trap cartridge, a two-position switching valve, and controlling computer software and hardware. The sample cleanup system was interfaced directly with the ESI source of a triple-stage-quadrupole MS using multiple reaction monitoring of negative product ions derived from S-PMA and the internal standard as the detection mode. The calibration curve was linear using human urine spiked at concentrations from 0.23 to 100 mg/L S-PMA (R2 = 0.997). The detection limit of the analytical system for neat S-PMA standard solution was 0.04 microg/L, whereas the detection limit was estimated to be lower than 0.35 microg/L for a urine matrix containing trace amounts of S-PMA. Without tedious manual sample cleanup procedures, the analytical system is fully automatic and therefore useful for high-throughput urinary S-PMA determination. With the selectivity and the sensitivity provided by MS-MS detection, the analytical system can be used for high-throughput and accurate determination of S-PMA levels in human urinary samples as a biomarker for benzene exposure.

Download full-text PDF

Source
http://dx.doi.org/10.1093/jat/26.4.205DOI Listing

Publication Analysis

Top Keywords

sample cleanup
16
cleanup system
12
benzene exposure
12
human urine
12
analytical system
12
online automatic
8
automatic sample
8
quantitative detection
8
detection benzene
8
exposure biomarker
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!