Optimal set-up for ultrasound guided punctures using new scanner applications: an in-vitro study.

Eur J Ultrasound

Department of Diagnostic Radiology, Roskilde Hospital, University of Copenhagen, Kogevej 7-13, DK 4000, Roskilde, Denmark.

Published: June 2002

Objective: To investigate if US-visualisation of the needle tip echo during US-guided punctures could be improved by use of new technological applications.

Method: an US-guided 18 G Trocar needle was inserted into targets of a puncture phantom. The punctures were performed in series of 10 using different settings of the US-scanner (GE Logic 700 Expert). At 7-cm of puncture depth the quality of the echo was tested using four different settings; normal (N), N + automatic tissue optimising (ATO), coded harmonic imaging (CHI), CHI + ATO and at 13-cm of puncture depths six different settings; N, N + ATO, coded excitation (CE), CE+ATO, CHI, CHI+ATO. In total 100 (40 + 60) images were randomly numbered and read independently by three radiologists with regard to scoring of the quality of the echoes from the needle tip, needle shaft and the target.

Results: US visualisation of the needle tip was significantly (P < 0.005) improved as compared to normal settings (N) when the settings of ATO and CE were used. CHI resulted in the lowest score. A high agreement between observers was registered. Similar results were registered with regard to scorings from the needle shaft and target.

Conclusion: Not only changes of needle designs and puncture techniques can enhance echoes from the needle but also changes in the settings of the US-scanner with the use of new technological applications can improve visualisation of the needle echo.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0929-8266(01)00170-7DOI Listing

Publication Analysis

Top Keywords

needle
9
needle echo
8
settings us-scanner
8
ato coded
8
settings ato
8
echoes needle
8
needle shaft
8
visualisation needle
8
settings
6
optimal set-up
4

Similar Publications

Purpose: This study aimed to investigate the usefulness of ultrasound-guided core-needle biopsy (US-CNB) for diagnosing type 1 AIP and evaluate the radiological outcomes following steroid therapy.

Materials And Methods: From January 2017 to June 2023, patients with pathology results containing "lymphoplasmacytic infiltration" and "fibrosis" were enrolled. The detection rate of level 1 histology by International Consensus Diagnostic Criteria (ICDC) and the contribution of US-CNB were assessed.

View Article and Find Full Text PDF

Matchbox Janus membrane fog collector with highly efficient directional transport.

Nanoscale Horiz

January 2025

Ministry of Education Key Laboratory for the Green Preparation and Application of Functional Materials, Hubei University, Wuhan 430062, People's Republic of China.

Coordinating the droplet capture, transport, and shedding processes during fog collection to achieve efficient fog collection is a major challenge. In this study, a copper mesh with different wettability was prepared by chemical etching and thiol modification. The Cu(OH) needle structure on the surface of the samples was characterized by FE-SEM and EDS tests, and the surface of the samples was chemically analyzed by infrared and XPS analyses.

View Article and Find Full Text PDF

Background: Telecytology-assisted rapid on-site evaluation (ROSE) offers a cost-effective method to enhance minimally invasive biopsies like fine needle aspiration and core biopsies with touch preparation. By reducing nondiagnostic sampling and the need for repeat procedures, ROSE via telecytology facilitates prompt triage for ancillary tests, improving patient management. This study examines cases initially deemed adequate for diagnosis during telecytology-assisted ROSE but later categorized as nondiagnostic at final evaluation (NDIS).

View Article and Find Full Text PDF

Therapeutic monoclonal antibody (mAb) drug products are increasingly used to treat both chronic and acute diseases. These mAb drug products are often developed for subcutaneous (SC) injection to simplify dosing compared with intravenous (IV) infusion. For SC injection, the mAb liquid drug product is typically filled in a vial for use with a syringe or in a prefilled syringe, which can then be assembled into a safety syringe device or an autoinjector for direct administration.

View Article and Find Full Text PDF

Rosai-Dorfman disease (RDD) is a self-limited, idiopathic, non-neoplastic disorder characterized by the proliferation of phagocytic histiocytes, which can mimic malignant lymphoproliferative disease. Cases of RDD most commonly present as bilateral painless cervical lymphadenopathy, with lesser involvement of the axilla, inguinal, and mediastinal lymph nodes. We present the case of a 62-year-old woman with a history of endometrial serous carcinoma who underwent evaluation at a dedicated breast imaging department after positron emission tomography/computed tomography (PET/CT) revealed breast masses and axillary nodes with increased uptake of fluorodeoxyglucose (FDG).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!