Comparison of Portapres with standard sphygmomanometry in pregnancy.

Hypertens Pregnancy

Department of Obstetrics, Academic Medical Center, P.O. Box 22660, 1100 DD Amsterdam, The Netherlands.

Published: June 2002

Background: Continuous beat-to-beat noninvasive blood pressure (BP) measurement is possible with Portapres. It constructs finger arterial waveforms beat-to-beat. Dedicated software is used to analyze the arterial waveforms. A new technique has been developed to reconstruct brachial intra-arterial pressure that uses return to flow (RTF). This method has been validated against invasive intra-arterial measurements in nonpregnant individuals.

Objectives: To validate Portapres in normal and preeclamptic pregnant women against standard aneroid sphygmomanometry according to Riva-Rocci-Korotkoff (RRK).

Methods: In 30 normotensive (10 in each trimester) and 20 preeclamptic women, two trained observers blinded from each other's results took BP measurements with a standard sphygmomanometer. These measurements were compared with sequential same-arm averaged measurements obtained during 30 sec by Portapres, following protocols from the Association for the Advancement of Medical Instrumentation (AAMI, mean accepted difference < or = 5 mmHg, SD < or = 8) and British Hypertension Society (BHS, gradings A down to D).

Results: A total of 150 measurement pairs were analyzed. Cumulative percentages of absolute pressure differences for all women (BHS) and mean pressure differences (SD) for different trimesters and preeclampsia (AAMI) between sphygmomanometry and Portapres were calculated. Overall, mean difference (SD) for systolic BP was 5 (SD 8) and for diastolic BP was -3 (SD 8), although analysis of variance revealed a significant effect for preeclampsia on diastolic differences between the two methods of BP measurement ( p<0.001).

Conclusions: Portapres with RTF, developed to equal intra-arterial brachial pressure, compares reasonably well to RRK and overall meets the criteria set by the AAMI. According to the BHS, Portapres receives a B-grading for diastolic BP and a C-grading for systolic BP. As Portapres measures BP and calculates cardiac output continuously and noninvasively, it would appear worthwhile to further evaluate this device in pathological pregnancies.

Download full-text PDF

Source
http://dx.doi.org/10.1081/PRG-120002910DOI Listing

Publication Analysis

Top Keywords

arterial waveforms
8
pressure differences
8
comparison portapres
4
portapres standard
4
standard sphygmomanometry
4
sphygmomanometry pregnancy
4
pregnancy background
4
background continuous
4
continuous beat-to-beat
4
beat-to-beat noninvasive
4

Similar Publications

The accurate non-invasive detection and estimation of central aortic pressure waveforms (CAPW) are crucial for reliable treatments of cardiovascular system diseases. But the accuracy and practicality of current estimation methods need to be improved. Our study combines a meta-learning neural network and a physics-driven method to accurately estimate CAPW based on personalized physiological indicators.

View Article and Find Full Text PDF

Flow-Based Coronary Artery Bypass Graft Patency Metrics: Uncertainty Quantification Simulations to Guide Development.

Cardiovasc Eng Technol

January 2025

Transonic Systems Inc., 34 Dutch Mill Road, Ithaca, New York, 14850, USA.

Purpose: Over time, transit time flow measurement (TTFM) has proven itself as a simple and effective tool for intra-operative evaluation of coronary artery bypass grafts (CABGs). However, metrics used to screen for possible technical error show considerable spread, preventing the definition of sharp cut-off values to distinguish between patent, questionable, and failed grafts. The simulation study presented in this paper aims to quantify this uncertainty for commonly used patency metrics, and to identify the most important physiological parameters influencing it.

View Article and Find Full Text PDF

Patients With .

J Neurotrauma

December 2024

Department of Medical Sciences, Section of Neurosurgery, Uppsala University, Uppsala, Sweden.

This study compared the roles of extraparenchymal autonomic nervous system (ANS) control of cerebral blood flow (CBF) versus intraparenchymal cerebrovascular autoregulation in 487 patients with aneurysmal subarachnoid hemorrhage (SAH) and 413 patients with traumatic brain injury (TBI). Vasomotion intensity of extraparenchymal and intraparenchymal vessels were quantified as the amplitude of oscillations of arterial blood pressure (ABP) and intracranial pressure (ICP) in the very low frequency range of 0.02-0.

View Article and Find Full Text PDF

: Aortic pulse wave velocity (aPWV) is a well-established surrogate marker of arterial stiffness. The Antares algorithm offers a method for determining aPWV from oscillometric blood pressure waveforms without requiring additional inputs. This prospective study aimed to evaluate the association and prognostic value of aPWV, determined by Antares, in predicting major adverse cardiovascular events (MACE).

View Article and Find Full Text PDF

Objective: To determine if the compensatory reserve algorithm validated in humans can be applied to canines. Our secondary objective was to determine if a simpler waveform analysis could predict the percentage of blood loss volume.

Methods: 6 purpose-bred, anesthetized dogs underwent 5 rounds of controlled hemorrhage and resuscitation while continuously recording invasive arterial blood pressure waveforms in this prospective, experimental study.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!