Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objective: To compare a custom-made femoral stem with a conventional cementless stem for initial stability in human femurs.
Design: The bone-femoral prosthesis interface motion was measured in vitro using displacement transducers. Combined axial and torsional loads simulating single-leg stance, loading were applied.
Background: The custom-made prosthesis is a cementless femoral stem fabricated from the patient's CT-reconstruction. To justify its clinical use its design has to be tested regarding his primary stability.
Methods: Seven pairs of human cadaveric femurs were used for testing the custom-made prosthesis versus a conventional cementless stem. Subsidence, rotation and interface motion were measured with load cycles up to 2000 N.
Results: The critical interface motion of 150 microm for the cementless prosthesis was not exceed for both types of prosthesis. The custom-made prosthesis enlarged in its metaphyseal part showed a decrease of dynamic micromotions and rotation behavior in the proximal region compared to the Alloclassic prosthesis stem.
Conclusions: The initial stability of the custom-made femoral stem was comparable to that of a conventional femoral stem.
Relevance: The data set for initial stability justifies the clinical use of a custom-made femoral stem in primary hip replacement. An individualized stem fit into the endosteal cortical bone decrease micromotion.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/s0268-0033(02)00012-8 | DOI Listing |