Paper strips inoculated with spores of Bacillus stearothermophilus ATCC 7953 were conventionally dried (lot 1) and lyophilized (lot 2); stored in defined environments of 32 and 86% relative humidity at 10, 25 and 33 degrees C for 210 d; and submitted to moist heat treatments at 121 degrees C. A significant decrease in thermal resistance from initial starting levels was found for lyophilized bioindicators stored at 86% relative humidity. The respective average D121 degrees C values were 1.55+/-0.05 and 1.37+/-0.10 min for lyophilized bioindicators stored at 32 and 86% relative humidity; and 1.65+/-0.15 min and 1.57+/-0.11 min for dried bioindicators stored in the same environments.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1385/abab:98-100:1-9:525 | DOI Listing |