We identified the error sources in a system for measuring tissue resistivity at eight frequencies from 1 Hz to 1 MHz using the four-terminal method. We expressed the measured resistivity with an analytical formula containing all error terms. We conducted practical error measurements with in-vivo and bench-top experiments. We averaged errors at all frequencies for all measurements. The standard deviations of error of the quantization error of the 8-bit digital oscilloscope with voltage averaging, the nonideality of the circuit, the in-vivo motion artifact and electrical interference combined to yield an error of +/- 1.19%. The dimension error in measuring the syringe tube for measuring the reference saline resistivity added +/- 1.32% error. The estimation of the working probe constant by interpolating a set of probe constants measured in reference saline solutions added +/- 0.48% error. The difference in the current magnitudes used during the probe calibration and that during the tissue resistivity measurement caused +/- 0.14% error. Variation of the electrode spacing, alignment, and electrode surface property due to the insertion of electrodes into the tissue caused +/- 0.61% error. We combined the above errors to yield an overall standard deviation error of the measured tissue resistivity of +/- 1.96%.

Download full-text PDF

Source
http://dx.doi.org/10.1109/10.995687DOI Listing

Publication Analysis

Top Keywords

tissue resistivity
16
error
13
resistivity measurement
8
reference saline
8
resistivity +/-
8
caused +/-
8
resistivity
6
+/-
6
tissue
5
error analysis
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!