Evaluation of intermittent pneumatic compression systems.

J Tissue Viability

School of Health Care Professions, University of Salford, Salford.

Published: April 2002

The operational characteristics of eight intermittent pneumatic compression systems were evaluated. Considerable differences between the pressure/time curves were found when different pumps were compared. The pressure attained in the garments varied from 6 to 124 mmHg and the rate of inflation from less than 1 to 9 mmHg per second.