Purpose: Important criteria for optimized stent implantation are high radial force and complete apposition of the stent itself. The aim of this study was to develop a simple method to assess these parameters under controlled experimental conditions and to compare vascular stents of different designs with regard to these parameters.

Materials And Methods: Five balloon-expandable stents of different designs (Jostent Flex, MAC stent, ML-Tristar, ML-Ultra, and S670) were tested. Fourteen stents of each type were implanted in a curved plastic vessel model (curve radius, 10 mm; lumen diameter, 3.5 mm) with use of a balloon inflation pressure of 12 atm. Part of the model was a nonflexible eccentric stenosis which was 2 mm or 8 mm in length. After stent implantation, a stenosis of 10%-70% was induced and radial force of the stent struts was measured at the site of the stenosis. The apposition of the stent to the vessel wall was imaged by high-resolution radiography.

Results: Analysis of variance showed significant differences of radial force between the tested stents (P <.001). The ML-Tristar and ML-Ultra stents had the highest radial force with maximum mean values of 687 cN and 846 cN at a stenosis of 70% in the model with the long stenosis. The radial force of the S670 stent was the lowest whereas the MAC stents showed an intermediate radial force. Radial force of the ML-Ultra stent was as much as 3.8 times higher than the force of the S670 stent (P <.001). Fifty percent of the ML-Tristar and ML-Ultra stents did not expand sufficiently to touch the vessel surface at the outer curvature. With an inflation pressure of 17 atm, complete apposition of these stents was achieved. The highest number of apposition irregularities was found in the S670 group (13 of 14 stents), whereas the MAC stents revealed the lowest number of irregular appositions (three of 14 stents). A significant correlation was found between the number of interconnecting struts and the number of irregular apposition events (P <.01).

Conclusions: This model allows an accurate in vitro evaluation of different stent parameters, such as apposition to the vessel wall and radial force. None of the investigated stents showed optimal results with respect to both parameters. The apposition behavior was significantly influenced by the architecture of the stents.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s1051-0443(07)61530-9DOI Listing

Publication Analysis

Top Keywords

radial force
16
vascular stents
8
vessel model
8
stent implantation
8
apposition stent
8
stents designs
8
stent
6
stents
5
radial
4
force wall
4

Similar Publications

An automated micro-tweezers system with a flexible workspace would benefit the intelligent sorting of live cells. Such micro-tweezers could employ a forced vortex strong enough to capture a single cell. Furthermore, addressable control of the position to the vortex would constitute a robotic system.

View Article and Find Full Text PDF

Objective: The use of vertical margin design in all-ceramic restoration has generated inquiries regarding its clinical efficacy under diverse dynamic oral conditions. This research aims to assess the marginal fit and fracture resistance of monolithic zirconia crowns featuring vertical margin design as opposed to those with conventional horizontal margin design.

Materials And Methods: Two metal dies were employed to generate replicated resin dies mimicking mandibular first molar preparation.

View Article and Find Full Text PDF

Purpose: This study aims to analyze microvascular reconstruction in Oral and Maxillofacial Surgery (OMFS) in Europe.

Methods: Based on previous studies, a dynamic online questionnaire was developed and subjected to internal and external evaluation. The questionnaire comprised multiple-choice, rating, and open-ended questions, addressing general and specific aspects and the impacts of the COVID-19 pandemic on microvascular reconstruction in OMFS in Europe.

View Article and Find Full Text PDF

Radially Distributed Electron Transfer on Single-Crystalline Surface of Gold Microplates.

ACS Nano

January 2025

State Key Laboratory for Modification of Chemical Fibers and Polymer Materials, College of Chemistry and Chemical Engineering, Donghua University, Shanghai 201620, China.

Electron transfer is ubiquitous in many chemical reactions and biological phenomena; however, the spatial heterogeneities of electron transfer kinetics in electrocatalysis are so far insufficiently resolved. Measuring and understanding the localized electron transfer are crucial to deciphering the intrinsic activity of electrocatalysts and to achieving further improvements in performance. By using scanning electrochemical probe microscopy to spatially resolve redox electrochemistry across the single-crystalline surface of gold microplates, we discover an intriguing radially distributed electron transfer pattern, where the kinetics around the periphery region are significantly higher than those at the central region, regardless of the redox reaction types.

View Article and Find Full Text PDF

Single atom alloys aggregation in the presence of ligands.

Nanoscale

January 2025

Department of Chemical and Petroleum Engineering, University of Pittsburgh, Pittsburgh, Pennsylvania 15261, USA.

Single atom alloys (SAAs) have gained tremendous attention as promising materials with unique physicochemical properties, particularly in catalysis. The stability of SAAs relies on the formation of a single active dopant on the surface of a metal host, quantified by the surface segregation and aggregation energy. Previous studies have investigated the surface segregation of non-ligated and ligated SAAs to reveal the driving forces underlying such phenomena.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!