Background: A hemostatic monitor capable of rapid, accurate detection of clinical coagulopathy within the operating room could improve management of bleeding after cardiopulmonary bypass (CPB). The Clot Signature Analyzer is a hemostatometer that measures global hemostasis in whole blood. The authors hypothesized that point-of-care hemostatometry could detect a clinical coagulopathic state in cardiac surgical patients.

Methods: Fifty-seven adult patients scheduled for a variety of elective cardiac surgical procedures were studied. Anesthesia, CPB, heparin anticoagulation, protamine reversal, and transfusion for post-CPB bleeding were all managed by standardized protocol. Clinical coagulopathy was defined by the need for platelet or fresh frozen plasma transfusion. The Clot Signature Analyzer collagen-induced thrombus formation (CITF) assay measured platelet-mediated hemostasis in vitro. The activated clotting time, platelet count, prothrombin time, activated partial thromboplastin time, and fibrinogen concentration were also measured.

Results: The postprotamine CITF was greater in patients who required hemostatic transfusion than in those who did not (17.6 +/- 8.0 min vs. 10.5 +/- 5.7 min, respectively; P < 0.01). Postprotamine CITF values were highly correlated with platelet and fresh frozen plasma transfusion (Spearman r = 0.50, P < 0.001 and r = 0.40, P < 0.005, respectively). Receiver operator characteristic curves showed a highly significant relation between the postprotamine CITF and intraoperative platelet and fresh frozen plasma transfusion (area under the curve, 0.78-0.81, P < 0.005) with 60-80% sensitivity, specificity, positive and negative predictive values at cutoffs of 12-14 min. Logistic regression demonstrated that the CITF was independently predictive of post-CPB hemostatic transfusion, but standard hemostatic assays were not.

Conclusions: The Clot Signature Analyzer CITF detects a clinical coagulopathic state after CPB and is independently predictive of the need for hemostatic transfusion. Hemostatometry has potential utility for monitoring hemostasis in cardiac surgery.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00000542-200205000-00014DOI Listing

Publication Analysis

Top Keywords

clot signature
16
signature analyzer
16
platelet fresh
12
fresh frozen
12
frozen plasma
12
plasma transfusion
12
postprotamine citf
12
hemostatic transfusion
12
hemostasis cardiac
8
cardiac surgery
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!