Background: Argon plasma coagulation has been rapidly accepted for endoscopic obliteration of vascular lesions and superficial tumors. Depth of injury is thought to be limited through preferential discharge of energy to nondesiccated tissue. However, argon plasma coagulation-induced injury has not been well characterized. The aim of this study was to characterize argon plasma coagulation-induced colonic injury by using a porcine model.

Methods: Laparotomy was performed in 6 female swine and the colon exteriorized with the subjects under general anesthesia. Lesions were made with an argon plasma coagulation probe held perpendicular and 2 mm from the mucosa. Variables studied were as follows: power (45 W, 60 W, and 75 W) and duration (1, 2, or 3 seconds; n = 11 for each power/duration combination). Injury was graded as either superficial or deep, involving the muscularis propria.

Results: Circular muscle layer injury correlated closely with power (p = 0.02), duration (p = 0.001), and total energy delivered (r = 0.977). Longitudinal muscle damage was associated with duration of burn (p = 0.001) and total energy delivered (r = 0.855), but correlated poorly with power (p = 0.40). No perforations occurred. Submucosal injection of saline solution had a protective effect with reductions in circular (90% to 10%, p = 0.002) and longitudinal muscle injury (50% to 0%, p = 0.1).

Conclusions: Injury to the muscularis propria occurs at recommended settings for argon plasma coagulation. Injury correlates with power setting, duration of burn, and total energy delivery. Protective arcing to nondesiccated tissue does not appear to be significant in vivo. Submucosal injection of saline solution protects against deep injury.

Download full-text PDF

Source
http://dx.doi.org/10.1067/mge.2002.123418DOI Listing

Publication Analysis

Top Keywords

argon plasma
24
plasma coagulation
16
total energy
12
injury
10
nondesiccated tissue
8
plasma coagulation-induced
8
0001 total
8
energy delivered
8
longitudinal muscle
8
duration burn
8

Similar Publications

High-Performance Edge-Contact Monolayer Molybdenum Disulfide Transistors.

Research (Wash D C)

January 2025

School of Integrated Circuits and Beijing Advanced Innovation Center for Integrated Circuits, Peking University, Beijing 100871, China.

Edge contact is essential for achieving the ultimate device pitch scaling of stacked nanosheet transistors with monolayer 2-dimensional (2D) channels. However, due to large edge-contact resistance between 2D channels and contact metal, there is currently a lack of high-performance edge-contact device technology for 2D material channels. Here, we report high-performance edge-contact monolayer molybdenum disulfide (MoS) field-effect transistors (FETs) utilizing well-controlled plasma etching techniques.

View Article and Find Full Text PDF

Introduction: Thermal ablative methods (such as argon plasma coagulation (APC) and soft tip snare coagulation (STSC) are commonly used to treat polyp margins. We aim to appraise the current literature and compare clinical outcomes between patients with treated (with APC vs. STSC) and non-treated endoscopic mucosal resection (EMR) margins.

View Article and Find Full Text PDF

Background And Aims: Obesity is a global health concern. Bariatric surgery offers reliably effective and durable weight loss and improvements of other comorbid conditions. However, the accessibility of bariatric surgery remains limited.

View Article and Find Full Text PDF

In this work, within the framework of a self-consistent model of arc discharge, a simulation of plasma parameters in a mixture of argon and methane was carried out, taking into account the evaporation of the electrode material in the case of a refractory and non-refractory cathode. It is shown that in the case of a refractory tungsten cathode, almost the same methane conversion rate is observed, leading to similar values in the density of the main methane conversion products (C, C, H) at different values of the discharge current density. However, with an increase in the current density, the evaporation rate of copper atoms from the anode increases, and a jump in the - characteristic is observed, caused by a change in the plasma-forming ion.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!