Assessing errors in the determination of base excess.

Anesth Analg

Department of Anesthesiology and Intensive Care Medicine, Kraukenaustalten Gilead, von Bodelschwinghsche Anstalten Bethel, Bielefeld, Germany.

Published: May 2002

Unlabelled: We compared estimates for base excess of extracellular fluid (BE(ecf); mmol/L) obtained in five clinically used blood gas analyzers: AVL Compact 2 (Roche Diagnostics, Mannheim, Germany), Ciba-Corning 860 (Bayer Diagnostics, Fernwald, Germany), IL 1620 (Instrumentation Laboratories, Lexington, MA), Stat Profile Ultra (Nova Biomedical, Waltham, MA), and ABL 510 (Radiometer, Copenhagen, Denmark). A total of 134 measurements per analyzer were obtained in arterial and venous blood samples from 10 patients undergoing cardiac surgery and 65 measurements per analyzer in venous blood samples from 2 healthy volunteers. The blood samples were equilibrated in a tonometer with gases of known composition (37 degrees C). Additional theoretical studies were performed to evaluate the relationship between pH and calculated BE(ecf) value (with varied PCO(2)) using the formulas of the various analyzers. The standard deviations of repeated measurements were 0.24 mmol/L for ABL 510 and approximately 0.45 mmol/L for the other 4 analyzers. The maximal systematic difference between the average of all measurements of each analyzer was 3.7 mmol/L; this was primarily attributable to differences in measuring pH, and, to a lesser extent, to differences in calculation and determination of PCO(2). Comparison of the results from samples with different oxygen saturation showed that the relative alkalinity of deoxygenated hemoglobin (Haldane effect) can also influence the determinations of BE(ecf).

Implications: A clinically useful way to quantify nonrespiratory disturbances of the acid-base balance is calculation of the base excess of extracellular fluid by using blood gas analyzers. In this study, we found significant variability in estimates of base excess of extracellular fluid obtained with five analyzers from different manufacturers. This variability is attributable to multiple factors, including lack of correction for deoxygenated hemoglobin (Haldane effect).

Download full-text PDF

Source
http://dx.doi.org/10.1097/00000539-200205000-00015DOI Listing

Publication Analysis

Top Keywords

base excess
16
excess extracellular
12
extracellular fluid
12
measurements analyzer
12
blood samples
12
estimates base
8
blood gas
8
gas analyzers
8
abl 510
8
venous blood
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!