Purpose: In the present study, the impact of the two different fat suppression techniques was investigated for free breathing 3D spiral coronary magnetic resonance angiography (MRA). As the coronary arteries are embedded in epicardial fat and are adjacent to myocardial tissue, magnetization preparation such as T(2)-preparation and fat suppression is essential for coronary discrimination.
Materials And Methods: Fat-signal suppression in three-dimensional (3D) thin- slab coronary MRA based on a spiral k-space data acquisition can either be achieved by signal pre-saturation using a spectrally selective inversion recovery pre-pulse or by spectral-spatial excitation. In the present study, the performance of the two different approaches was studied in healthy subjects.
Results: No significant objective or subjective difference was found between the two fat suppression approaches.
Conclusion: Spectral pre-saturation seems preferred for coronary MRA applications due to the ease of implementation and the shorter cardiac acquisition window.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/jmri.10078 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!