Purpose: The current bedside "gold standard" for cardiac output (CO) monitoring is thermodilution using a pulmonary artery catheter (PAC) but there is a number of risks associated with its use. The primary objective of this review was to evaluate the utility of esophageal Doppler (ED) as a minimally invasive monitor of CO.

Source: Medline literature search from 1966 to 2001 with citation review for studies comparing ED to PAC thermodilution for CO in perioperative and critically ill patients.

Principal Findings: Twenty-five publications were identified comparing ED and PAC measurement of CO in a broad range of patients. There was a good overall correlation between CO determined by ED and thermodilution (n = 18 studies, median R = 0.89, range 0.52 to 0.98) and minimal bias (n = 13, median -0.01, range 1.38 to 2 L x min(-1)). The precision of ED was only fair overall as assessed by limits of agreement. The ED technique was found to be responsive in detecting changes in thermodilution CO and was reliable demonstrating both low intra- and inter-observer variation. ED was reportedly easy to insert after minimal training and was safe, with no significant complications identified.

Conclusion: ED is a practical, reliable, and valid device for measuring CO in perioperative and critically ill patients. Further studies with larger numbers of patients are needed to determine if the limited precision observed is inherent to the technique, the diagnoses of patients studied, or the small sample sizes.

Download full-text PDF

Source
http://dx.doi.org/10.1007/BF03017329DOI Listing

Publication Analysis

Top Keywords

utility esophageal
8
esophageal doppler
8
doppler minimally
8
minimally invasive
8
comparing pac
8
perioperative critically
8
critically ill
8
invasive hemodynamic
4
hemodynamic monitor
4
monitor review
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!