An ion trap/ion mobility/quadrupole/collision cell/time-of-flight mass spectrometer that incorporates a differentially pumped orifice-skimmer cone region at the back of the drift tube has been developed for the analysis of peptide mixtures. The combined approach allows a variety of strategies to be employed for collisionally activating ions, and fragments can be monitored by subsequent stages of mass spectrometry in a parallel fashion, as described previously (Anal. Chem. 2000, 72, 2737). Here, we describe the overall experimental approach in detail. Applications involving different aspects of the initial mobility separation and various collisional activation and parallel sequencing strategies are illustrated by examining several simple peptide mixtures and a mixture of tryptic peptides from beta-casein. Detection limits associated with various experimental configurations and the utility for analysis of complex systems are discussed.

Download full-text PDF

Source
http://dx.doi.org/10.1021/ac010837sDOI Listing

Publication Analysis

Top Keywords

peptide mixtures
12
collisional activation
8
analysis complex
8
coupling ion
4
ion mobility
4
mobility separations
4
separations collisional
4
activation techniques
4
techniques multiple
4
multiple stages
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!