Unlabelled: Newer pulse oximeters have been developed to be motion resistant and thus have few false alarms. However, they have not yet been evaluated in a pediatric sleep laboratory setting. While evaluating new oximeters for use in our laboratory, we obtained simultaneous pulse oximetry data from two Masimo oximeters and from two Nellcor oximeters during nocturnal polysomnography in children referred for sleep-disordered breathing (SDB). In series 1, comprising 24 patients, comparisons were made between a Masimo oximeter with 4-second averaging time and the Nellcor N-200 oximeter set for 3 to 5 second averaging. A maximum of 20 events per patient were randomly selected for analysis, an "event" being a desaturation of > or = 4% registered by either oximeter. Interobserver agreement for event classification was 93%. Eighty-eight percent of 220 desaturation events occurring during wakefulness and 38% of 194 events occurring during sleep were classified as motion artifact on the Nellcor oximeter. Neither the Masimo oximeter nor the transcutaneous oxygen probe confirmed that the desaturation was real, in most of these cases. During sleep, there were 119 events detected by either or both oximeters: 113 (95%) by the Nellcor versus 82 (69%) by the Masimo. For these 119 events, the extent of desaturation was slightly less for the Masimo than the Nellcor oximeter, 4.5 +/- 2.4% versus 5.5 +/- 2.5%, respectively. In series 2, 22 patients were studied comparing a Masimo Radical oximeter with 2 second averaging to the Nellcor N-200 oximeter. The extent of desaturation was slightly greater for the Masimo oximeter. The Masimo oximeter detected more non-artifactual desaturation events occurring during sleep than the Nellcor oximeter, 90% versus 76% (chi2 = 9.9, p < 0.01). In series 3, comprising 128 events in 5 patients, a Nellcor N-395 oximeter detected fewer desaturations during non-movement, sleep periods and had more movement related "desaturation" events, compared to a Masimo Radical oximeter.

Conclusions: The Masimo oximeters register many fewer false desaturations due to motion artifact. Using 4-second averaging, a Masimo oximeter detected significantly fewer SaO2 dips than the Nellcor N-200 oximeter but using 2-second averaging, the Masimo oximeter detected more SaO2 dips than the Nellcor N-200 oximeter. The sensitivity and motion artifact rejection characteristics of the Nellcor N-395 oximeter are not adequate for a pediatric sleep laboratory setting. These findings suggest that in a pediatric sleep laboratory, use of a Masimo oximeter with very short averaging time could significantly reduce workload and improve reliability of desaturation detection.

Download full-text PDF

Source

Publication Analysis

Top Keywords

masimo oximeter
28
oximeter
18
nellcor n-200
16
n-200 oximeter
16
oximeter detected
16
masimo
13
pediatric sleep
12
sleep laboratory
12
events occurring
12
motion artifact
12

Similar Publications

Objectives: To determine oxygen saturation (SpO) values and their variability in healthy-term neonates at high altitude.

Methods: In a prospective multisite, non-randomized blinded study, we performed SpO measurements in healthy neonates admitted to postnatal wards and followed up in Public Health Nurse Clinics (PHNC) in Calgary, Alberta, Canada. We recruited healthy-term neonates (≥37 weeks) between 6 and 24 h of age.

View Article and Find Full Text PDF

Motion resistance in peripheral oxygen saturation monitoring using Biolight Analog SpO compared to Masimo SpO: a non-inferiority study.

BMC Anesthesiol

November 2024

Department of Clinical Applications, Guangdong Biolight Meditech Co., Ltd., No.2 Innovation 1st Road, Technical Innovation Coast, Hi-tech Zone, Zhuhai, Guangdong, 519085, China.

Background: Pulse oximeters are vital for assessing blood oxygen levels but can produce inaccurate readings during patient motion, leading to false alarms and alarm fatigue. Analog SpO Technology, which uses analog waveforms to filter motion artifacts, may improve accuracy compared to digital sensors. However, the effectiveness of this technology in reducing false alarms in clinical settings remains unclear.

View Article and Find Full Text PDF

This study evaluates the suitability of cerebral oximetry using near-infrared spectroscopy (NIRS) compared to traditional pulse oximetry (SpO2) for measuring cerebral oxygen saturation during hypoxia testing, aiming to enhance safety during flight operations and training. The study included 106 participants aged 18-60 years at the Aerospace Medicine Training Center in Königsbrück. Cerebral oxygen saturation (rSO2) and peripheral oxygen saturation (SpO2) were measured using the INVOS™ 5100C cerebral oximeter and Masimo™ MS5 pulse oximeter, respectively.

View Article and Find Full Text PDF

Background: The data is evolving on exhaled carbon dioxide (ECO) levels for preterm infants requiring stabilisation.

Objectives: To establish the trends of ECO levels during the first 10 minutes of stabilisation in preterm infants at birth.

Methods: We conducted a multi-centre, prospective observational study.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!