Objective: The study investigated the hypothesis that threshold and comfortable levels recorded from cochlear implant patients would reduce, and dynamic range increase, as distance of the electrode from the modiolar wall (radial distance) decreases. Two groups of cochlear implant patients participated; one group using the Nucleus' 24 Contour electrode array, and one group using the Nucleus standard straight (banded) array. The Nucleus 24 Contour array has been shown in temporal bone studies to lie closer to the modiolus than the banded array. The relationship of electrode impedance and radial distance is also investigated.

Design: The study, conducted at three centers, evaluated 21 patients using the Contour array, and 36 patients using the banded array. For each patient, threshold, comfortable levels and dynamic range were measured at four time points. Common ground electrode impedance was recorded clinically from each patient, at time intervals up to 12 wk. An estimate of the radial distance of the electrode from the modiolus was made by analysis of Cochlear view x-rays.

Results: Threshold and comfortable levels were significantly lower for the Nucleus 24 Contour array than for the banded array. However, dynamic range measurements did not show the predicted increase. In a majority of subjects, a significant correlation was found between the estimated radial distance of the electrode from the modiolus and the measured threshold and comfortable levels. This trend was not observed for dynamic range. The analysis indicates that other factors than radial distance are involved in the resultant psychophysical levels. Clinical impedance measures (common ground) were found to be significantly higher for the Contour array. However, the electrodes on the Contour array are half-rings, which are approximately only half the geometric size of the full rings as electrodes of the standard array. When the geometric electrode area in the two array designs are normalized, the trends in the electrode impedance behavior are similar.

Conclusions: The results support the hypothesis that the relationship between the radial distance of the electrode and the psychophysical measures are influenced by patterns of fibrous tissue growth and individual patient differences, such as etiology and neural survival. Impedance measures for the Nucleus 24 Contour electrode array were higher than the banded electrode array, but this is primarily due to the reduction in electrode surface area. The different outcomes in impedance over time suggest differences in the relative contributions of the components of impedance with the two arrays.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00003446-200202001-00004DOI Listing

Publication Analysis

Top Keywords

radial distance
24
threshold comfortable
20
contour array
20
comfortable levels
16
dynamic range
16
distance electrode
16
banded array
16
array
14
electrode
12
electrode array
12

Similar Publications

Cephalometric analysis is the primary diagnosis method in orthodontics. In our previous study, the algorithm was developed to estimate cephalometric landmarks from lateral facial photographs of patients with normal occlusion. This study evaluates the estimation accuracy by the algorithm trained on a dataset of 2320 patients with added malocclusion patients and the analysis values.

View Article and Find Full Text PDF

The human upper limb has undergone various evolutionary myologic changes, accompanied by corresponding modifications in the anatomical course of neurovascular structures. In this study, we aimed to elucidate the emergence of the accessory head of the flexor pollicis longus (AHFPL) muscle as a beneficial biomechanical evolutionary development and its topographical relationship with adjacent neurovascular structures. In this pursuit to understand this phenomenon, dissections were conducted on sixty-two upper limbs from thirty-one cadavers.

View Article and Find Full Text PDF

Background: We assessed the accuracy and performed a directional analysis of robot-assisted implantation of stereoelectroencephalography (SEEG) depth electrodes in children using the frameless neurolocate 3D registration module.

Methods: Thirteen children with epilepsy undergoing stereotactic robot-assisted insertion of SEEG electrodes were included. Six children were operated on with standard frame-based registration while 7 with the use of the frameless neurolocate registration module.

View Article and Find Full Text PDF

Purpose: To evaluate visual and refractive outcomes in eyes with a history of radial keratotomy (RK) implanted with the second-generation light-adjustable lens (LAL).

Setting: Private practice, multiple locations.

Design: Retrospective, consecutive case series.

View Article and Find Full Text PDF

Cilia and flagella play a crucial role in the development and function of eukaryotes. The activity of thousands of dyneins is precisely regulated to generate flagellar motility. The complex proteome (600+ proteins) and architecture of the structural core of flagella, the axoneme, have made it challenging to dissect the functions of the different complexes, like the regulatory machinery.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!