  • A clinical study evaluated hemolysis in 278 patients with either mechanical or bioprosthetic heart valves to understand its frequency and severity.
  • Results showed that while none of the patients developed anemia, mild subclinical hemolysis was present in 49 patients, predominantly among those with mechanical valves.
  • Factors such as double valve replacement and the type of valve (mechanical vs bioprosthetic) significantly influenced the presence and severity of hemolysis.

Article Abstract

Objective: A prospective clinical study was designed to assess the frequency and severity of intravascular hemolysis in patients with new-generation, normally functioning prosthetic heart valves.

Methods: Hemolysis was evaluated in 172 patients with a mechanical prosthesis (53 CarboMedics and 119 Sorin Bicarbon) and in 106 patients with a bioprosthesis (15 St Jude Medical Toronto, 19 Baxter Perimount, and 72 Medtronic Mosaic) in the aortic position, mitral position, or both. Aortic valve replacement was performed in 206 patients, mitral valve replacement in 59 patients, and double valve replacement in 13 patients. The presence of hemolysis was assessed on the basis of the level of serum lactic dehydrogenase and serum haptoglobin and the presence and amount of reticulocytes and schistocytes in the peripheral blood. Severity of intravascular hemolysis was estimated on the basis of serum lactic dehydrogenase. Clinical, echocardiographic, and hematologic evaluations were performed 1, 6, and 12 months after discharge.

Results: None of the 278 patients experienced decompensated anemia, whereas at 12 months, mild subclinical hemolysis was identified in 49 patients, 44 (26%) with a mechanical prosthesis and 5 (5%) with a bioprosthesis (P <.001). At multivariate analysis, independent predictors of the presence of subclinical hemolysis were mitral valve replacement (P <.001), use of a mechanical prosthesis (P =.002), and double valve replacement (P =.02). Frequency of hemolysis in patients with stented aortic bioprostheses was 3%, whereas it was absent in those with stentless valves. Among mechanical valve recipients, double versus single valve replacement (P =.04) and mitral versus aortic valve replacement (P =.05) were correlated with the presence of hemolysis; double valve recipients also showed a more severe degree of hemolysis (P =.03). In patients with a Sorin Bicarbon prosthesis, hemolysis was less frequent (22% vs 34%, P =.09) and severe (P <.001) than in those with a CarboMedics prosthesis.

Conclusions: In normally functioning prosthetic heart valves, subclinical hemolysis is a frequent finding. A low incidence of hemolysis is found in stented biologic prostheses, and it is absent in stentless aortic valves. Modifications of valve design may contribute to minimize the occurrence of hemolysis in mechanical prostheses.

Download full-text PDF

Source
http://dx.doi.org/10.1067/mtc.2002.120337DOI Listing

Publication Analysis

Top Keywords

intravascular hemolysis
12
valve replacement
12
patients
9
hemolysis patients
8
patients new-generation
8
prosthetic heart
8
severity intravascular
8
mechanical prosthesis
8
replacement patients
8
serum lactic
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!