OBJECTIVE: To evaluate two rapid immunoassay methods, QuickVue-Chlamydia (Quidel Corp., San Diego California) and Kodak SureCell (Kodak Corp., Rochester, NY) for the detection of Chlamydia trachomatis antigen in endocervical swabs from high- and low-risk females. METHODS: Seven hundred and twenty-four females attending three clinics were enrolled in the study. The results were compared to McCoy's or BGMK cell culture and discrepancies resolved with polymerase chain reaction and direct fluorescent antibody tests performed on left-over culture specimens. RESULTS: The sensitivity, specificity, predictive value of a positive and predictive value of a negative of the QuickVue Chlamydia assay were 92.0%, 99.1%, 92.0% and 99.1%, respectively. The sensitivity, specificity, predictive value of a positive and predictive value of a negative of the SureCell assay were 90.0%, 99.8%, 98.6% and 98.8%, respectively. CONCLUSIONS: The performances of the two immunoassay methods were similar, and slight differences in sensitivity and specificity were not statistically significant. Both immunoassay methods performed well in high- and low-risk patient groups, both for symptomatic and for asymptomatic patients.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1469-0691.1997.tb00475.xDOI Listing

Publication Analysis

Top Keywords

immunoassay methods
16
sensitivity specificity
12
rapid immunoassay
8
detection chlamydia
8
chlamydia trachomatis
8
trachomatis antigen
8
antigen endocervical
8
high- low-risk
8
specificity predictive
8
predictive positive
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!