Article Abstract

Aims: To establish a reliable method for measuring the axial length of the eye with incomplete filling of silicone oil in the vitreous cavity in order to determine the IOL power before executing combined cataract surgery and silicone oil removal.

Methods: 12 eyes of 12 patients undergoing combined cataract extraction and silicone oil removal between October 1998 and June 2000 were entered prospectively into this study. All eyes were examined using an x ray computed tomography (CT) scanner. Each axial length of a silicone oil injected eye was measured on the best CT slice among eyeball cuts, which showed both the insertions of the medial and lateral rectus muscle, the thickest slice of the lens, and the optic nerve. The IOL power for the eyes was determined using the SRK/T formula based on the CT data. The deviation of postoperative refraction from the goal refraction was evaluated for each eye.

Results: The CT slices demonstrated that the vitreous cavity was not completely filled with the oil in all of the silicone oil injected eyes. The deviation of postoperative refraction from the goal refraction was less than 1 dioptre (D) error in six of 12 eyes (50%), and was less than 2 D error in nine eyes (75%). In the other three eyes having more than 2 D error, the axial length was 27 mm or more.

Conclusion: The CT measurement of axial eye length for determining IOL might be a useful method for evaluating silicone oil injected eyes before combined cataract surgery and silicone oil removal.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1770965PMC
http://dx.doi.org/10.1136/bjo.86.1.47DOI Listing

Publication Analysis

Top Keywords

silicone oil
32
axial length
16
vitreous cavity
12
combined cataract
12
oil injected
12
eyes
9
oil
9
measurement axial
8
incomplete filling
8
silicone
8

Similar Publications

Engineering Wettability Transitions on Laser-Textured Shark Skin-Inspired Surfaces via Chemical Post-Processing Techniques.

Micromachines (Basel)

November 2024

Queensland Micro- and Nanotechnology Centre, Griffith University, Nathan Campus, 170 Kessels Road, Brisbane, QLD 4111, Australia.

Surface wettability, the interaction between a liquid droplet and the surface it contacts, plays a key role in influencing droplet behavior and flow dynamics. There is a growing interest in designing surfaces with tailored wetting properties across diverse applications. Advanced fabrication techniques that create surfaces with unique wettability offer significant innovation potential.

View Article and Find Full Text PDF

Ocular hypertension after silicone oil filling surgery for high myopia: a case-control study.

BMC Surg

January 2025

Chongqing Nanping Aier Ophthalmology, 249 Nancheng Avenue, Nanan District, Chongqing, P. R. China.

Purpose: This study aims to examine the correlation between specific clinical parameters, such as axial eye length, and the onset of ocular hypertension "OH" following the use of silicone oil filling in patients with high myopia.

Method: In this retrospective analysis, we reviewed 214 eyes from 432 patients diagnosed with severe myopia, all of whom underwent vitrectomy and were treated with silicone oil filling. The study aimed to document the incidence and timing of postoperative ocular hypertension "OH" while assessing various factors, including demographic details, medical history, additional surgical interventions, and findings from supplementary examinations (such as axial length, silicone oil emulsification, and anterior chamber penetration).

View Article and Find Full Text PDF

Multifunctional aluminum alloy slippery liquid-infused surface with porous and boehmite nanoflower structure.

J Colloid Interface Sci

December 2024

Ministry of Education Key Laboratory for the Green Preparation and Application of Functional Materials, Hubei University, Wuhan 430062, People's Republic of China; State Key Laboratory of Solid Lubrication, Lanzhou Institute of Chemical Physics, Chinese Academy of Sciences, Lanzhou 730000, People's Republic of China. Electronic address:

The depletion of lubricants in (slippery liquid-infused porous surfaces) SLIPS poses a significant challenge to their long-term functionality. While line-shaped rough structures can mitigate lubricant loss to some extent, they often fail to provide the stability required for sustained performance. In this study, we present a novel porous nanoflower aluminum alloy slippery liquid-infused surface (P-NF-AA SLIPS), which integrates a porous framework with a rough nanoflower structure.

View Article and Find Full Text PDF

Purpose: The emulsification of silicone oil (SO) remains poorly understood. In the present study, we investigated the physical properties of unused pharmaceutical SO samples under various conditions. Moreover, clinical correlations with the patients' SO samples were assessed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!