Objective: The goal of this study was to assess effects of translation (horizontal displacement) and angulation (transducer tilting) on 3-dimensional (3D) echocardiographic volumes of both balloons and human left ventricles after autopsy.

Methods: Six water-filled (non-) aneurysmatic balloons of 150, 250, and 350 mL and 3 hearts of different sizes and shapes were suspended upright in a water bath. Angulation and/or translation was performed respectively by tilting the transducer with a mechanical arm in a vertical plane relative to the balloon tip or true apex of the hearts and by shifting the water bath in the same vertical plane. For balloon and left ventricular (LV) volume assessment, a 3D conical data set was obtained by TomTec rotational acquisition in combination with a HP Sonos 5500 ultrasound machine.

Results: For the 6 balloons, translation from 1 to 4 cm yielded volumes of up to 74% of the optimal volume (100%); angulation of 10 degrees or 20 degrees, volumes of up to 80% and 34%. Translation with 10-degree angulation yielded volumes up to 64%; for 20-degree angulation and translation, there was no volume loss. Results were similar for the left ventricles.

Conclusions: Even minor angulation or translation of the transducer yields substantial underestimation of the true volume. Off-axis para-apical views, however, defined as angulation of 20 degrees and greater than 0.5 cm translation in this in vitro model, obviate volume underestimation. Such views in patients, if obtainable, may be an attractive alternative for conventional apical 3D acquisition, especially in dilated and aneurysmatic hearts.

Download full-text PDF

Source
http://dx.doi.org/10.1067/mje.2002.117293DOI Listing

Publication Analysis

Top Keywords

volume assessment
8
rotational acquisition
8
water bath
8
vertical plane
8
yielded volumes
8
angulation degrees
8
angulation translation
8
translation
7
angulation
7
volume
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!