Introduction: Motion artifact and low perfusion often lead to faulty or absent pulse oximetry readings in clinical practice.
Objective: Determine the impact of motion artifact and low perfusion on newly introduced pulse oximetry technologies during hypoxemic episodes in healthy volunteers.
Methods: Five different pulse oximeters from 4 manufacturers (the Datex Ohmeda 3900P; the Agilent; the Nellcor N-3000; the Nellcor N-395; and the Schiller OX-1, which is the European version of the Ivy SatGuard 2000 with Masimo SET) were compared with respect to their ability (separated or in combination) to provide accurate readings in the presence of motion artifact and low perfusion. Four of these oximeters represent the latest available oximetry technology, and one (the N-3000) represents a previous generation of oximeters. Oxygen saturation values (S(pO(2))) and pulse rate from the oximeters were recorded during episodes of induced hypoxemia in 10 healthy volunteers. Standardized and repeatable motion artifacts were generated by a motion machine and by having the test subject perform tapping and scratching motions. Perfusion to the finger was reduced by an inflatable balloon impinging on the brachial artery. The pulse oximetry readings from the test oximeters were compared to readings from control pulse oximeters on the unperturbed reference hand. The pulse rates from the test oximeters were compared to the electrocardiographically-measured heart rate.
Results: The frequency of faulty readings was increased by increasing motion interference and decreasing perfusion. The S(pO(2)) deviation was within +/- 3% of the reference reading > 95% of the time for all instruments during the control desaturation period in the absence of motion and with normal perfusion. With the combination of motion and low perfusion, the S(pO(2)) error was within +/- 3% less than 62% of the time for all oximeters tested. A significant difference in the frequency of large S(pO(2)) errors was observed only in the direct comparison of the N-395 and N-3000. The N-395 exhibited less frequent S(pO(2)) error exceeding 6% of S(pO(2)) in the combination of the most challenging situations (motion and motion with reduced perfusion). In the same situation the Datex-Ohmeda 3900P and Nellcor N-3000 showed significantly higher pulse rate errors than the other devices (Datex-Ohmeda 3900P 53% of the time and N-3000 37% of the time).
Conclusions: The established model of creating motion artifact and low perfusion is capable of simulating a hierarchy of severe clinical situations. With solely motion or solely reduced perfusion the percentage of errors exceeding +/- 3% of S(pO(2)) increased by 20% and 10%, respectively, compared to the control period. Simultaneous presence of motion and reduced perfusion leads to a relative incidence of > 35% of errors > 3% of S(pO(2)) for the various oximeters. In this situation the N-3000 and the Datex-Ohmeda 3900P exhibited differences between estimated pulse rate and electrocardiographically-measured heart rate > 25 beats/min > 37% of the time.
Download full-text PDF |
Source |
---|
J Clin Med
December 2024
Physio-Pathology and Immunology Department, "Carol Davila" University of Medicine and Pharmacy, 050474 Bucharest, Romania.
: With the development of artificial intelligence (A.I.), the optical coherence tomography angiography (OCTA) analysis of progression in hypertensive retinopathy could be improved.
View Article and Find Full Text PDFChildren (Basel)
November 2024
Guangdong Cardiovascular Institute, Guangdong Provincial People's Hospital, Guangdong Academy of Medical Sciences, Guangzhou 510080, China.
Objectives: Our goal was to evaluate the associations between postoperative coronary perfusion pressure (CPP) values and in-hospital outcomes in infants after congenital cardiac surgery. Our goal was to assess the relationship between postoperative coronary perfusion pressure (CPP) values and in-hospital outcomes in infants following congenital cardiac surgery.
Methods: In this study, we conducted a retrospective analysis on a cohort of 296 consecutive infant patients (aged 31-120 days) who underwent congenital cardiac surgery between 1 January 2019 and 30 April 2019.
Xenotransplantation (XTx) is an increasingly realistic solution to the organ shortage. Clinical XTx may require off-site procurement in a designated pathogen free (DPF) facility necessitating a period of cold ischemic time during transportation. This study evaluates the impact of different kidney preservation strategies on early graft function in pig-to-baboon XTx in a series of eight cases of pig-to-baboon xenotransplantation performed after five hours of cold ischemic time and compares these results to six cases of pig-to-baboon xenotransplantation performed with minimal ischemic time.
View Article and Find Full Text PDFJ Cereb Blood Flow Metab
January 2025
AP-HP, Hôpital Lariboisière, Department of Anaesthesia and Critical Care, Paris, France.
In patients with acute brain injury (ABI), optimizing cerebral perfusion parameters relies on multimodal monitoring. This include data from systemic monitoring-mean arterial pressure (MAP), arterial carbon dioxide tension (PaCO), arterial oxygen saturation (SaO), hemoglobin levels (Hb), and temperature-as well as neurological monitoring-intracranial pressure (ICP), cerebral perfusion pressure (CPP), and transcranial Doppler (TCD) velocities. We hypothesized that these parameters alone were not sufficient to assess the risk of cerebral ischemia.
View Article and Find Full Text PDFJ Korean Med Sci
January 2025
Division of Cardiology, Department of Internal Medicine, Chonnam National University Medical School, Gwangju, Korea.
Background: The ionic mechanism underlying Brugada syndrome (BrS) arises from an imbalance in transient outward current flow between the epicardium and endocardium. Previous studies report that artemisinin, originally derived from a Chinese herb for antimalarial use, inhibits the Ito current in canines. In a prior study, we showed the antiarrhythmic effects of artemisinin in BrS wedge preparation models.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!