A fast, sensitive and specific method is presented for the quantification of RSD921 in human plasma by liquid chromatography coupled with tandem mass spectrometry using tri-deuterated RSD921 (3d-RSD921) as an internal standard. A single-step liquid/liquid extraction was performed with diethyl ether/hexane (80 : 20, v/v) using 0.5 ml of plasma. The plasma calibration curves were linear from 0.1 to 20 ng ml(-1) (r > 0.999). Between-run precision, based on the percent relative deviation for replicate (n = 40) quality controls, was < or =7.27% (0.5 ng ml(-1)), < or =7.39% (5.0 ng ml(-1)), and < or =5.06% (20.0 ng ml(-1)). Between-run accuracies, based on the relative error, were +/-2.59%, +/-1.23% and +/-1.64% respectively. The method was developed to evaluate the pharmacokinetic profile after 15 min of intravenous stepwise-ascending infusion dose of RSD921 in 18 healthy volunteers. A dissociation study of protonated RSD921 and 3d-RSD921 by collision-induced dissociation using in-source fragmentation and tandem mass spectrometry is also presented.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/jms.218 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!