The paper, which is a continuation of the previous one, describes a six-channel, PC-dedicated implanted telemetry system, including the schematic drawings, and explanation of all the differences between a single channel and multichannel system approach. The scheme of an additional multichannel analog output, that makes a pure analog recording possible, even without the PC connection, is also presented. Specific methods for both parallel (Centronics) and serial (RS232C) port interfacing are demonstrated and the controlling software principles are discussed. The representative recordings obtained from animal experiments of simultaneous (direct and integrated) multichannel intestinal EMG (MMC) are shown. The device can also be used for the ECG, EEG and uterine smooth muscle electrical activity recording.