Background: Body heat loss during anaesthesia may result in increased morbidity, particularly in high-risk populations such as children. To avoid hypothermia, a novel thermoregulatory system (Allon) was devised. We tested the safety and efficacy of this system in maintaining normothermia in children undergoing routine surgical procedures.
Methods: The system consists of a computerized body, which receives continuous afferent data, i.e. core (rectal) temperature. These data are then compared with a preset temperature (37 degrees C) and a microprocessor heating/cooling unit warms/cools the temperature of circulating water in a garment that is specially designed to allow maximal coverage of body surface area, without impingement on the surgical field. Water temperature to the garment was limited to a maximum of 39.5 degrees C. Continuous perioperative monitoring of skin and rectal temperature, heart rate and blood pressure was performed. Postoperative shivering and adverse effects were also assessed.
Results: The Allon system was used in 38 patients aged 3 months to 14 years undergoing surgery under general anaesthesia lasting more than 30 min. Fifty to 80% body surface area was covered by the garment. Mean operative and postoperative core temperatures were 36.9 +/- 0.5 degrees C and 36.7 +/- 0.5 degrees C, respectively. Intraoperative skin temperatures were maintained at 34.4 +/- 2.7 degrees C. The average core- to-periphery intraoperative gradient was 2.9 +/- 4.9 degrees C. Postoperative shivering was absent in 36 cases and mild in two cases. No device-related adverse effects were observed.
Conclusions: Perioperative thermoregulation using the Allon system is safe and effective in maintaining body temperature within a narrow range in children undergoing brief surgical procedures.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1046/j.1460-9592.2001.00713.x | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!