Ability of new heart rate monitors to measure normal and abnormal heart rate.

J Sports Med Phys Fitness

Laboratoire Performance Motrice, UFR STAPS, Université Blaise Pasal, Aubière, Cedex, France.

Published: December 2001

AI Article Synopsis

  • The study aimed to evaluate the accuracy of two new heart rate monitors (Accurex Plus and Vantage NV) compared to an older model (PE 4000) to assess their ability to detect heart rate anomalies in non-sportive and older individuals.
  • In laboratory tests under normal and abnormal heart rate scenarios, the newer devices outperformed the older model, displaying significantly better accuracy during stable heart rate stages, especially in transitional phases when rates were changing.
  • The results showed that both Accurex Plus and Vantage NV had over 90% accuracy, while PE 4000 only reached 33%, indicating that newer monitors are more reliable for detecting heart rate problems.

Article Abstract

Background: The extended use of heart rate monitors to non sportive people, and older ones increases the risk to be facing heart rate troubles. Questions exist upon the ability of this devices to detect such abnormalities. The purpose of the investigation was to evaluate the accuracy of two third generation heart rate monitors, Accurex Plus and Vantage NV and to compared these data with those of an older one, PE 4000. Then we investigate responses of this monitors to abnormal heart rate.

Experimental Design: The three heart rate monitors were tested in the laboratory, wired in parallel to an ECG simulator under two modes: a normal heart rate program (7 stable heart rate stages with two marked transitions) and an automatic arrhythmia one.

Measures: 1st program: Values generated by the ECG simulator were compared to the values produced by the three devices. Particular attention was paid to the transition phases (heart rate drop and heart rate fall). Results of heart rate monitors accuracy were expressed at the exact value (devices value=simulator value) and at approximated value of +/-3 beats x min(-1) (devices value=simulator value +/-3 beats x min(-1)); 2nd program: Abnormal rhythms were analysed using an ambulatory ECG recorder (Synésis) as control data, and compared to the data from the three devices.

Statistics: Correlation between simulator generated true values and HRMs read values were calculated.

Results: In the 1st normal heart rate program, with an accuracy of exact value, Accurex Plus and Vantage NV, were more accurate than the PE 4000 (94% and 89% of values respectively versus 33% for the PE4000). At +/-3 beats x min(-1), the three devices gave good results: over 98% of total values. In transition phases: the three devices showed a smoothing effect, which was stronger in decelerating heart rate than in accelerating heart rate. In the 2nd program: isolated heart rhythm troubles (missing beat, pause <4 sec, supra-ventricular ectopic activity, and ventricular ectopic activity) were either not detected or were rejected. Signal changes such as changes in QRS (ventricular bigeminy) may cause detection losses, and so modify calculated heart rate. The HR signal rising/dropping slopes caused by sudden heart rhythm disorders, were notably attenuated, with a consequent loss of HR accuracy.

Conclusions: In laboratory stable heart rate conditions, third generation HRMs are more accurate than earlier ones. Heart rate monitors are less accurate in transient phases and have not been improved in that domain. The three HRMs ignore isolated heart rate troubles. As expected, the usefulness of HRMs in detecting HR disorders is limited. Nevertheless, these devices have some value with serious troubles, like pause, bradycardia or tachycardia, lasting longer than 4 sec, and especially if they coincide with functional symptoms. The use of such even more accurate devices may consequently be recommended to healthy public for which they were built.

Download full-text PDF

Source

Publication Analysis

Top Keywords

heart rate
56
rate monitors
20
heart
15
rate
14
three devices
12
+/-3 beats
12
beats min-1
12
abnormal heart
8
accurex vantage
8
compared data
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!