Breast compression and radiation dose in two different mammographic oblique projections: 45 and 60 degrees.

Eur J Radiol

Department of Diagnostic and Interventional Radiology, University Hospital Merkur, Zajceva 19, 10000 Zagreb, Croatia.

Published: October 2001

Introduction: Standard mammography includes two views, craniocaudal and medio-lateral oblique. Depending on patient's body constitution, central beam angle in mediolateral oblique projection may vary, with 45 degrees being suitable for the majority of patients in routine daily practice. With continuous improvement in X-ray technology and radiographers' training, the risk of radiation induced cancerogenesis is considerably reduced and acceptable when compared to benefit. However, the risk still exists, being cumulative and directly related to absorbed glandular dose. There is no minimal dose of radiation which is absolutely harmless, and every effort to reduce the dose is welcome. In this retrospective study two different angles (45 vs. 60 degrees) of mediolateral oblique view were compared according to radiation dose and efficacy of breast compression.

Patients And Methods: In 52 women, additional 60 degrees oblique films were done after craniocaudal and mediolateral oblique 45 degrees-films, with the same kVp and positioning technique. Breast thickness, time-current products (mA s) and absorbed doses were compared between 45 degrees- and 60 degrees-films. Subgroups of women with large, small, prominent and pendulous breasts were analyzed separately, following the same methodology as for the whole group.

Results: mA s were 11.5% lower and compression 7% better with an angle of 60 degrees than with 45 degrees. In the subgroup of women with small breasts, mA s values were 13% lower and compression 9% better with 60 degrees than with 45 degrees, while in the subgroup with large breasts, mA s were 9% lower and compression 5% better. In the subgroup of patients with pendulous breasts, mA s values were 12% lower and compression 10% better with 60 degrees than with 45 degrees, while in the subgroup with prominent breasts, mA s values were 4% lower and compression 3% better. Absorbed glandular dose was estimated to be approximately 20% lower when an oblique mammogram was done with 60 degrees instead of 45 degrees. The compression with 15 kp was well tolerated by the majority of patients.

Discussion And Conclusion: Mammograms of excellent quality should be done with as low a radiation dose as possible. Adequate breast compression is fundamental in mammography due to immobilization of the breast, shortening of the exposure times, reduction of motion and geometric blur and prevention of overpenetration by means of equalizing breast thickness. As the absorbed glandular dose cannot be accurately measured, it is convenient to estimate the dose approximately, on the basis of its linear proportionality with exposure dose. With constant technical properties of X-ray machines, exposure dose is determined only by mA s. Hence, the absorbed glandular dose in our study was influenced only by changes of mA s and breast thickness. As the absorbed dose reduction is proportional to the product of the reduction of mA s and thickness, we estimated that absorbed dose was 7-22% lower if 60 degrees is applied instead of 45 degrees. Breast compression and mA s were more favourable in women with pendulous breasts, possibly because of elongation of the glandular disc in the lateroascending direction, with its longer axis directed more perpendicularly. Fibroglandular tissue in the 60 degrees-view is thus projected onto a larger film area, with less effect of superimposition. In conclusion, because of lower mA s values and better compression, which finally result in a 25% lower absorbed dose, we recommend the oblique view be done with an angle of 60 degrees, especially for small and pendulous breasts.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0720-048x(01)00317-5DOI Listing

Publication Analysis

Top Keywords

lower compression
20
absorbed glandular
16
glandular dose
16
pendulous breasts
16
compression better
16
degrees degrees
16
dose
15
degrees
15
breast compression
12
radiation dose
12

Similar Publications

Ornithopod dinosaurs appeared during the Middle Jurassic, but it was in the Lower Cretaceous they started their successful evolutionary history. Different phylogenies describing the evolutionary relationships of Ornithopoda are mostly based on cranial features, however there is a lack of well-preserved and complete skulls for the basal member of the clade, hampering our knowledge on the mode and tempo of these herbivorous dinosaurs. Here we describe YLSNHM 01942, a well-preserved skull of a juvenile neornithischian from the Liaoning Province of China.

View Article and Find Full Text PDF

The primary weight-bearing structure of the proximal femur, trabecular bone, has a complex three-dimensional architecture that was previously difficult to comprehensively display. This study examined the spatial architecture of trabecular struts in the coronal, sagittal, and horizontal sections of the proximal femur using 21 cases prepared with P45 sectional plasticization. The primary compressive strut (PCS) exhibited a "mushroom-like" shape with upper and lower parts.

View Article and Find Full Text PDF

Analysis of the Feasibility of the OrthoNail Hybrid Intramedullary Implant in the Human Body with Respect to Material Durability.

J Funct Biomater

January 2025

Department of Mechanics, Materials and Biomedical Engineering, Faculty of Mechanical Engineering, Wrocław University of Science and Technology, Smoluchowskiego 25, 50-372 Wrocław, Poland.

This study focuses on the development and evaluation of the OrthoNail hybrid intramedullary implant for lower limb lengthening in patients requiring significant skeletal reconstruction. The implant addresses the challenges in load-bearing during rehabilitation, providing a robust solution that is capable of supporting physiological loads. Mechanical tests, including axial compression, tension, torsion, and 3,4-point bending, determined the implant's load capacity and fatigue resistance, while finite element analysis assessed stress distributions in bone tissue and around screw holes during single-leg stance, with boundary conditions derived from Orthoload database data.

View Article and Find Full Text PDF

The Effect of Lipopolysaccharides from on the Size, Density, and Compressibility of Phospholipid Vesicles.

Biomimetics (Basel)

January 2025

Department of Nuclear Physics and Biophysics, Faculty of Mathematics, Physics and Informatics, Comenius University, Mlynska Dolina F1, 84248 Bratislava, Slovakia.

The properties of the large unilamellar vesicles (LUVs) from 1,2-dimyristoyl-sn-glycero-3-phosphatidylcholine (DMPC), modified by lipopolysaccharides (LPS) from sv. Enteritidis, which mimics Gram-negative bacteria, were studied by various physical methods. LPS, in the range of 0/20/50 % / relative to the lipid, had a regulatory role in the structure of the LUVs toward the lower size, low polydispersity, and over-a-month size stability due to the lower negative zeta potential.

View Article and Find Full Text PDF

With increasing mining depth, the coal pillars of a coal mine will be in a stressful environment characterized by high gas pressures and unidirectional loading. To investigate the damage evolution characteristics and energy evolution mechanism of coal pillars loaded in a gas pressure environment, a uniaxial compression test was performed on a coal body under different gas pressures using a load testing apparatus for gas-containing coal rocks. The obtained results showed that the mechanical properties of the coal body varied with the gas pressure.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!