Ventriculocoronary artery bypass results using a mesh-tipped device in a porcine model.

Ann Thorac Surg

Cardiac Surgical Associates, Minneapolis, Minnesota 55407, USA.

Published: September 2001

Background: In this report we describe the in vivo evaluation of a device and ventriculocoronary artery bypass procedure that creates a permanent transmyocardial channel between the left ventricle and a coronary artery.

Methods: The transmyocardial device, an L-shaped titanium tube with a meshed distal tip and an exterior polyester cuff, was implanted from the base of the left ventricle to the proximal left anterior descending coronary artery in 11 healthy juvenile domestic pigs using a beating-heart approach. Flow rates were measured at implant. Patency was assessed at explant for surviving animals at 2 (n = 3) and 4 weeks (n = 4).

Results: Flow through the transmyocardial device after implantation was 74% of base line. Forward flow occurred during systole. Luminal patency was 100% at 2 weeks and 75% at 4 weeks. Histologic analysis showed little to no intimal proliferation at the coronary interface.

Conclusions: This short-duration study shows promise for perfusing ischemic myocardium with systolic flow. The transmyocardial titanium conduit and treated coronary artery patency was good at 2 and 4 weeks and warrants further studies.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0003-4975(01)02947-2DOI Listing

Publication Analysis

Top Keywords

ventriculocoronary artery
8
artery bypass
8
left ventricle
8
transmyocardial device
8
coronary artery
8
flow transmyocardial
8
bypass mesh-tipped
4
device
4
mesh-tipped device
4
device porcine
4