Purpose: The present study was conducted to evaluate 4 different fluid delivery systems: Arthrex AR-6450, Stryker 1.5L High Flow Pump, Arthro FMS 4, and Acufex InteliJet. Their flow and pressure capacity, as specified by the manufacturer, and their accuracy, as measured in a standardized fashion, were compared with directly measured pressure values.

Type Of Study: Experimental study.

Methods: Two experiments were performed: (1) the achieved pressure was measured with the pressure sensor within a ball and compared with the preset pressure values. The pressure was increased constantly while the flow was kept constant. (2) maximum flow per minute was tested using a beaker and a stopwatch.

Results: The highest measured pressures for the Arthrex and the Stryker pumps were 115 mm Hg (46% of published maximum pressure) and 113 mm Hg (57% of published maximum pressure), respectively. The highest measured pressures for the Arthro FMS 4 and the InteliJet pumps were 251 mm Hg (101% of published pressure) and 132 mm Hg (88% of published maximum pressure), respectively. The maximal flow values for the Arthrex and Stryker pumps were 900 mL/min (56.3%) and 675 mL/min (48%), respectively. The Arthro FMS 4 and InteliJet pumps reached a maximal flow of 450 mL/min (56%) and 1,250 mL/min (62.5%).

Conclusions: In the simpler pumps (Arthrex AR-6450 and Stryker 1.5L High Flow Pump), the pressure on the display must be set to higher values compared with the outflow-controlling pumps (Arthro FMS 4 and InteliJet) to achieve the same intra-articular pressure. All pumps were able to maintain a pressure of 60 mm Hg very accurately. Therefore, a surgeon can trust all of the pumps when the pressure is set below 60 mm Hg.

Download full-text PDF

Source
http://dx.doi.org/10.1053/jars.2001.25340DOI Listing

Publication Analysis

Top Keywords

arthro fms
16
pressure
14
published maximum
12
maximum pressure
12
fms intelijet
12
flow
8
arthrex ar-6450
8
ar-6450 stryker
8
stryker 15l
8
15l high
8

Similar Publications

Purpose: The present study was conducted to evaluate 4 different fluid delivery systems: Arthrex AR-6450, Stryker 1.5L High Flow Pump, Arthro FMS 4, and Acufex InteliJet. Their flow and pressure capacity, as specified by the manufacturer, and their accuracy, as measured in a standardized fashion, were compared with directly measured pressure values.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!