Background: The haemodynamic monitor PiCCO System, based on transpulmonary arterial thermodilution, has been used with a brachial-axillary access instead of the femoral arterial access during abdominal aortic aneurysm surgical repair. Accuracy and limitations of pulse contour continuous cardiac output (PCCO) were evaluated on the basis of arterial thermodilution cardiac output. The patterns of cardiac index, preload, afterload and cardiac function parameters were also studied in the different phases of the surgical procedure.
Methods: Twenty consecutive patients were studied. Mean differences (bias) between PCCO and arterial thermodilution cardiac output were calculated by the Bland-Altman test. Analysis of variance with multiple comparison test of haemodynamic variables in the different phases were performed. The correlation coefficients between cardiac index and the volumetric preload variables were also obtained.
Results: Brachial artery catheterization was achieved without any major complication. Pulse contour continuous cardiac index (CI) and arterial thermodilution CI values showed overall mean differences (bias) of -0.04 Lámin-1. m-2 (SD 0.8) but after aortic cross-clamping and aortic unclamping they were 0.64 Lámin-1. m-2 (SD 0.57) e -0.57 Lámin-1. m-2 (SD 0.85), respectively (p<0.05). CI, global end-diastolic volume (GEDV) and intrathoracic blood volume (ITBVI) were significantly lower during aortic cross-clamping. CI was not correlated to central venous pressure (r=0.18) but instead, to GEDV (r=0.57) and ITBVI (r=0.65).
Conclusions: PiCCO System with brachial-axillary arterial access was suitable for haemodynamic monitoring of the abdominal aortic aneurysm surgical repair procedures. PCCO must be recalibrated with arterial thermodilution after aortic cross-clamping and unclamping to avoid an over-estimation and an under-estimation respectively. During aortic cross-clamping GEDV and ITBVI indicated a decreased preload. Other haemodynamic variables were less valuable but EVLWI showed an interesting steady increase during the whole procedure.
Download full-text PDF |
Source |
---|
Life (Basel)
January 2025
National Pulmonary Hypertension Service, Hammersmith Hospital, Imperial College NHS Trust, DuCane Road, London W12 0HS, UK.
Background: Stroke volume index (SVI) is an important prognostic parameter in pulmonary arterial hypertension (PAH). The direct Fick (DF) method represents the gold standard for measuring it. Indirect Fick (IF) and thermodilution (TD) are simpler and widely used alternatives.
View Article and Find Full Text PDFJ Cardiovasc Med (Hagerstown)
February 2025
Center of Excellence in Cardiovascular Sciences, Ospedale Isola Tiberina, Gemelli Isola.
Aims: Coronary microvascular dysfunction (CMD) is a heterogeneous condition defined by reduced coronary flow reserve (CFR). The new index 'microvascular resistance reserve' (MRR) has been developed, but its role is unclear. We investigate the relationships between functional indices in ANOCA (angina and non-obstructive coronary arteries) patients and evaluate the hemodynamic features of different CMD subtypes.
View Article and Find Full Text PDFLife Sci
January 2025
Department of Cardiology, Aarhus University Hospital, Palle Juul-Jensens Boulevard 69, 8200 Aarhus N, Denmark; Department of Clinical Medicine, Aarhus University Hospital, Palle Juul-Jensens Boulevard 11, 8200 Aaarhus N, Denmark.
Background: Butyrate, a short-chain fatty acid, has shown potential to improve left ventricular (LV) function and induce vasorelaxation in rodents. Butyrate may either be produced by the microbiome in the colon, be ingested or administered intravenously. This study aimed to evaluate effects of butyrate on cardiac output (CO) and associated hemodynamic variables in a porcine model.
View Article and Find Full Text PDFAnn Intensive Care
January 2025
Department of Intensive Care, CHIREC Hospitals, Université Libre de Bruxelles, Brussels, Belgium.
Int J Cardiol
January 2025
Center of Excellence in Cardiovascular Sciences, Ospedale Isola Tiberina, Gemelli Isola, Via di Ponte Quattro Capi, 39, 00186 Rome, Italy; Department of Cardiovascular and Pneumological Sciences, Università Cattolica del Sacro Cuore, Largo Agostino Gemelli, 8, 00168 Rome, Italy. Electronic address:
Aims: To investigate the influence of index of microcirculatory resistance (IMR) on fractional flow reserve (FFR) and adenosine-induced hyperemia (ΔPd/Pa-FFR) in patients with chronic (CCS) or stabilized acute coronary syndromes (ACS), utilizing various IMR threshold values.
Methods: Data were extracted from two ongoing Italian registries involving patients with CCS or stabilized ACS who underwent a #FullPhysiology approach [Pd/Pa, FFR, IMR, coronary flow reserve (CFR)] by bolus thermodilution technique in the left anterior descending artery. Correlations between IMR and both FFR and ΔPd/Pa-FFR were analyzed both globally and within three IMR-defined groups: Group 1 (IMR <25), Group 2 (25 ≤ IMR <40), and Group 3 (IMR ≥40).