Background: Uniplanar quantitative angiography (QA) is the standard method for measuring vessel diameter during surgical and endovascular procedures. Intravascular ultrasound (IVUS), a relatively new technology, is another means of obtaining this measurement. This study was designed to validate the accuracy of these two modalities by comparing each to direct caliper measurement, the gold standard, using phantom femoral artery segments (PAS).

Materials And Methods: PAS diameter was measured with a 12.5-MHz mechanically rotating IVUS catheter (Boston Scientific Corp.) and QA (OEC Corp.) was compared to the direct caliper measurement (Mitutoyo Corp.) at 60 different locations within PAS. At each location minimal lumen diameter and perpendicular lumen diameter were measured and their mean was calculated. The intraclass correlation coefficients (ICCC) between direct caliper measurement and IVUS and uniplanar and biplanar angiography were calculated. Fisher's Z transformation was used to compare the correlation coefficients.

Results: The ICCC for IVUS was 0.89. The ICCCs for uniplanar and biplanar angiography were 0.73 and 0.82, respectively. IVUS correlated more closely with direct caliper measurement than uniplanar and biplanar angiography (P = 0.00008, 0.02) Biplanar angiography correlated more closely with direct caliper measurement than uniplanar angiography (P = 0.04).

Conclusions: IVUS more accurately measures lumen diameter than uniplanar or biplanar angiography. Diameter measurement with biplanar angiography is more accurate than uniplanar angiography.

Download full-text PDF

Source
http://dx.doi.org/10.1006/jsre.2001.6214DOI Listing

Publication Analysis

Top Keywords

biplanar angiography
24
direct caliper
20
caliper measurement
20
uniplanar biplanar
16
lumen diameter
12
angiography
9
intravascular ultrasound
8
measurement
8
diameter measurement
8
diameter measured
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!