Background: Congenital hearing disorder, which has an occurrence of 1 in 1000 newborn babies, is a health problem that should be taken seriously. Once it has been discovered too late, it leads to lifelong language, intellectual, emotional and social problems. The aim of this project was the critical evaluation of a universal, ABR-based newborn hearing screening, looking at the specificity and the success of the method used as well as the organisational requirements and costs involved with this method.

Methods: From August 1999 until March 2000, 464 newborn babies, born at the University Hospital Eppendorf in Hamburg, were examined with the Algo portable. Another 31 children, all younger than 6 months, were also examined with the Algo portable, because of suspected hearing loss. When a child got a refer twice by the Algo portable, the child was examined again, using the conventional ERA-machine (Evoselect) to confirm the suspected hearing loss and to assess its grade.

Results: 98% of the newborns at our hospital were examined. Up to now, out of 13 conspicuous children, the majority of which belonged to the risk group, 6 have been identified as unambiguously bilateral and 2 as unilateral hard of hearing. After an introductory period of two months, the average costs per examination could be set at 38.65 DM and app. 3200 DM per bilateral hearing loss discovered.

Conclusion: A Universal Newborn Hearing Screening is necessary and can be executed with calculable effort with concern to personnel, time and finance.

Download full-text PDF

Source
http://dx.doi.org/10.1055/s-2001-15708DOI Listing

Publication Analysis

Top Keywords

algo portable
16
newborn hearing
12
hearing screening
12
hearing loss
12
universal newborn
8
hearing
8
newborn babies
8
examined algo
8
suspected hearing
8
newborn
5

Similar Publications

Objective: Neonatal hearing screening programs in Flanders and the Netherlands use Natus ALGO screening devices. Since 2006 in Flanders and 2009 in the Netherlands, both programs have replaced the older ALGO Portable devices with the newer ALGO 3i devices. However, in both countries, ALGO 3i devices have a significantly higher rate of referrals than ALGO Portable devices.

View Article and Find Full Text PDF

MB11 BERAphone) hearing screening compared to ALGOportable in a Dutch NICU: a pilot study.

Int J Pediatr Otorhinolaryngol

October 2010

Department of Neonatology, Isala Clinics, Zwolle, The Netherlands.

Objective: Since 2001 all Dutch Neonatal Intensive Care Units use a highly reliable automated auditory brainstem response (AABR) newborn hearing screening device, the (ALGOportable). A more recent AABR newborn hearing screening device, the MB11 BERAphone, does not make use of disposable ear couplers and has a different algorithm, a possible advantage with regard to a faster response acquisition. Use of the MB11 BERAphone is therefore potentially more cost effective and faster.

View Article and Find Full Text PDF

Objective: Since 1998 an integrated universal newborn hearing screening programme (UNHSP) based on automated auditory brainstem response (AABR) has been implemented in Flanders. The protocol of the UNHSP is based on guidelines defined by the American Academy of Paediatrics (AAP). The aim of this paper is to report on the screening protocol and to assess its feasibility.

View Article and Find Full Text PDF

The aim of this study was to compare the Natus-ALGO Portable (Portable) device with the Natus-ALGO 2e and to evaluate its usefulness. Portable, an automated ABR screener (Natus Medical, Foster City, CA, USA) was used to screen 826 neonates in our hospital between February 2002 and March 2003. The referral rate for the first automated ABR screening was 0.

View Article and Find Full Text PDF

Background: Congenital hearing disorder, which has an occurrence of 1 in 1000 newborn babies, is a health problem that should be taken seriously. Once it has been discovered too late, it leads to lifelong language, intellectual, emotional and social problems. The aim of this project was the critical evaluation of a universal, ABR-based newborn hearing screening, looking at the specificity and the success of the method used as well as the organisational requirements and costs involved with this method.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!