Despite several studies showing a higher incidence of peri-implant femoral fractures with the Gamma nail than with a sliding screw plate (SSP), the Gamma nail has remained the standard implant for trochanteric fractures in many hospitals. We recorded 921 trochanteric fractures in the city of Oslo during 2 years and compared the reoperation frequency in patients treated with the Gamma nail (n 379) and SSP (n 542). The distribution of age and gender in the two treatment groups was the same. 65 patients were reoperated on, several of them more than once. The only significant difference between the two surgical methods in complications leading to a reoperation was the frequency of femoral shaft fractures. 17 of the patients treated with the Gamma nail had a new femoral fracture postoperatively, compared to 3 of those with a SSP. The relative risk of another femoral fracture after surgery was 12 (95 % CI: 2.7-52) if the surgical device was a Gamma nail compared to a SSP. The Gamma nail therefore can not be recommended as the standard implant for trochanteric fractures.

Download full-text PDF

Source
http://dx.doi.org/10.1080/00016470152846574DOI Listing

Publication Analysis

Top Keywords

gamma nail
28
trochanteric fractures
16
femoral fractures
8
fractures gamma
8
nail sliding
8
sliding screw
8
screw plate
8
ssp gamma
8
standard implant
8
implant trochanteric
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!