Hyperpolarized (129)Xe dissolved in a lipid emulsion constitutes an NMR tracer that can be injected into the blood stream, enabling blood-flow measurement and perfusion imaging. A small volume (0.15 ml) of this tracer was injected in 1.5 s in rat carotid and (129)Xe MR spectra and images were acquired at 2.35 T to evaluate the potential of this approach for cerebral studies. Xenon spectra consistently showed two resonances, at 194.5 ppm and 199.0 ppm relative to the gas peak. The signal-to-noise ratio (SNR) obtained for the two peaks was sufficient (ranging from 12 to 90) to follow their time courses. 2D transverse-projection xenon images were obtained with an in-plane resolution of 900 microm per pixel (SNR range 8-15). Histological analysis revealed no brain damage except in two rats that had received three injections.