Objective: To assess the accuracy of aortic valve area (AVA) calculations using the continuity equation with data obtained from the double envelope (DE) (simultaneously obtained left ventricular outflow tract [V1]) and aortic valve [V2] velocities) during intraoperative transesophageal echocardiography (TEE).

Design: Prospective study; measurements were performed on-line.

Setting: University hospital.

Participants: Cardiac and noncardiac surgical patients (n = 75) with recent aortic valve assessment (<3 months) undergoing general anesthesia or endotracheal intubation.

Interventions: Intraoperative AVA was measured by the continuity equation using the DE technique (DE/TEE) and by planimetry (PL/TEE). Left ventricular outflow tract diameter was obtained from midesophageal views, whereas subvalvular (V1) and valvular (V2) velocities were obtained simultaneously using continuous-wave Doppler from transgastric views. V1 was also obtained using pulsed-wave Doppler. Measurements were compared with AVA obtained preoperatively by the Gorlin equation during cardiac catheterization (G/CATH) or by transthoracic echocardiography using the traditional continuity equation (C/TTE) (nonsimultaneously obtained V1 and V2).

Measurements And Main Results: A DE was obtained in 73 of 75 patients (97%). Four patients had atrial fibrillation at the time of the examination, whereas the rest were in sinus rhythm. PL/TEE was performed in 54 of 71 patients with sinus rhythm (76%). Agreement was good between DE/TEE and G/CATH (mean bias, 0.02 cm(2) [SD, 0.24 cm(2)]), and C/TTE (mean bias, -0.05 cm(2) [SD, 0.16 cm(2)]). Agreement was not as good between PL/TEE and G/CATH (mean bias, -0.07 cm(2) [SD, 0.28 cm(2)]) and C/TTE (mean bias, -0.13 cm(2) [SD, 0.30 cm(2)]). V1 obtained by pulsed-wave Doppler and with DE closely agreed (mean bias, 0.01 m/sec [SD, 0.05 m/sec]).

Conclusion: TEE evaluation of native AVA using the DE technique is feasible and in good agreement with that obtained by C/TTE and G/CATH. Compared with DE/TEE, PL/TEE did not agree as well. Use of DE/TEE should simplify the continuity equation and may minimize errors resulting from beat-to-beat variability in stroke volume.

Download full-text PDF

Source
http://dx.doi.org/10.1053/jcan.2001.23272DOI Listing

Publication Analysis

Top Keywords

aortic valve
16
valve area
8
transesophageal echocardiographic
4
echocardiographic evaluation
4
evaluation native
4
aortic
4
native aortic
4
valve
4
area utility
4
utility double-envelope
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!