Introduction: The use of cannulated screw without protrusive head in Colles fractures could avoid some of the drawbacks of conventional pinning. In order to compare the mechanical resistance of Kirshner wires versus cannulated screws we designed a animal model of Colles fractures and tested three types of osteosynthesis: A: K-wires, B: Herbert cannulated screw, C: specific cannulated screws.

Method: After creating a 10 mm defect in three sets of 10 fresh turkey tibia, 3 types of osteosynthesis were done and tested in compression with an Adamel Lhomargy machine: set A: K-wire fixation, set B: Herbert screws fixation and set C: specifics screws fixation.

Results: The compression strength needed for failure of the fixations were: for the K-wires (set A): 52 N +/- 17; for the Herbert screws (set B): 93 N +/- 39; for the specifics screws (set C): 160 N +/- 48; (p < 0.0001).

Discussion: The use of an animal model makes the experimentation easier and the sampling more homogeneous. In this model, resistance to compression of the cannulated screw was better than K-wires and the specific cannulated screw better than Herbert screw. Therefore clinical trial of osteosynthesis with cannulated screw in Colles fracture could be considered.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s1297-3203(01)00031-2DOI Listing

Publication Analysis

Top Keywords

cannulated screw
20
cannulated
8
osteosynthesis cannulated
8
cannulated screws
8
colles fractures
8
animal model
8
types osteosynthesis
8
specific cannulated
8
fixation set
8
herbert screws
8

Similar Publications

Analysis of Extreme Beak Calcaneal Fracture (Type 2-Lee) Fixed with Cannulated Cancellous Screws: An Original Research Article.

J Orthop Case Rep

January 2025

Department of Orthopaedics, Vinayaka Mission's Medical College and Hospital, Vinayaka Mission's Research Foundation (DU), Karaikal , Puducherry, India.609609.

Introduction: Extreme beak calcaneal fractures (Type 2 Lee's tuberosity avulsion fractures) are rare injuries, accounting for only 1.3-3% of all calcaneal fractures. These injuries are considered as surgical emergency as they can lead to significant functional impairment and soft-tissue compromise if not promptly managed.

View Article and Find Full Text PDF

Introduction: Posterior cruciate ligament (PCL) avulsion fractures of the tibia with ipsilateral tibial shaft represent a rare but challenging orthopedic injury, necessitating careful consideration of surgical interventions for optimal outcomes. This case report presents the successful management of tibial shaft fracture (proximal 1/3rd junction) along with ipsilateral PCL avulsion fracture of tibia using a novel approach using open reduction internal fixation (ORIF) of tibial shaft with Locking compression plate (LCP) with cannulated cancellous screw and spiked washer fixation of PCL avulsion. There are no case reports or research articles available for the management of PCL avulsion fracture of tibia associated with ipsilateral tibial shaft fracture.

View Article and Find Full Text PDF

A 3D-printed guide is an effective method for accurately placing pedicle screws in dog vertebrae. While a conventional drill guide allows precise pilot hole formation, it can lead to potential screw wobbling during insertion. In this study, we applied a technique that assists with both drilling and screw insertion, and we compared the accuracy of screw placement using this approach with that achieved by the conventional drill guide.

View Article and Find Full Text PDF

Purpose: This study aims to describe a fixation technique for coronoid fractures using suture buttons, and to biomechanically evaluate this technique in comparison to screw fixation as a time-zero pilot study.

Methods: An O'Driscoll type 2 anteromedial coronoid facet (AMCF) fracture was simulated in 20 fresh-frozen human elbows. The specimens were randomized into two groups and fracture fixation was performed with either a suture button system or a 3.

View Article and Find Full Text PDF

Pediatric elbow fractures are quite common, accounting for up to 34% of bone fractures in children. Among these, acute lateral humeral condyle (LHC) fractures represent up to 22%. The accurate diagnosis and early treatment of LHC fractures are crucial due to the potential for abnormal growth and significant long-term impacts on joint motion.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!