Unlabelled: The purpose of this study was to investigate the quantitative properties and effects of ordered-subset expectation maximization (OSEM) on kinetic modeling compared with filtered backprojection (FBP) in dynamic PET studies. Both phantom and patient studies were performed.

Methods: For phantom studies dynamic two-dimensional emission scans with 10-min frames and 20-min scan intervals were acquired over a 14-h period using an HR+ PET scanner. Various phantoms were scanned: 2-, 5-, 10-, and 20-cm-diameter phantoms filled with an 18F solution (300 kBq/mL) and a NEMA phantom filled with an 18F background (40 kBq/mL) and a cold or 11C insert (450 kBq/mL). Transmission (Tx) scans of 5-60 min were acquired. Data were reconstructed using FBP Hanning 0.5 and OSEM with 2-12 iterations and 12 or 24 subsets. Quantitative accuracy and noise characteristics were assessed. For patient studies, five cardiac, three oncologic, and three brain dynamic 18F-FDG scans were used. Five reconstructions were performed: FBP Hanning 0.5, and OSEM 2 x 12 and OSEM 4 x 16 with and without 5-mm full width at half maximum smoothing. Time-activity curves were calculated using volumes of interest. The input function was derived from arterial sampling. Metabolic rate of glucose (MRglu) was calculated with a standard two-tissue compartment model and Patlak analysis.

Results: Contribution of Tx noise to the reconstructed image was smaller for OSEM than for FBP. Differences in signal-to-noise ratio between FBP and OSEM depended on number of iterations and phantom size. Bias with OSEM was observed for regions enclosed within a 5- to 10-fold hotter background. For cardiac studies OSEM 2 x 12 and OSEM 4 x 16 resulted in 13% and 21% higher pixel values and 9% and 15% higher MRglu values compared with FBP. Smoothing decreased all these values to 2%. Similar results were found for most tumor studies. For brain studies MRglu of FBP and OSEM 4 x 16 agreed within 2%. Use of OSEM image-derived input functions for cardiac PET studies resulted in a decrease in calculated MRglu of about 15%.

Conclusion: For most PET studies OSEM has equal quantitative accuracy as FBP. The higher pixel and MRglu values are explained by the better resolution of OSEM. However, OSEM does not provide accurate image-derived input functions for FDG cardiac PET studies because of bias in regions located within a hotter background.

Download full-text PDF

Source

Publication Analysis

Top Keywords

pet studies
16
osem
15
osem osem
12
studies
10
dynamic pet
8
kinetic modeling
8
fbp
8
patient studies
8
filled 18f
8
fbp hanning
8

Similar Publications

Purpose: Trophoblast cell-surface antigen 2 (Trop2) is overexpressed in various solid tumors and contributes to tumor progression, while its expression remains low in normal tissues. Trop2-targeting antibody-drug conjugate (ADC), sacituzumab govitecan-hziy (Trodelvy), has shown efficacy in targeting this antigen. Leveraging the enhanced specificity of ADCs, we conducted the first immunoPET imaging study of Trop2 expression in gastric cancer (GC) and triple-negative breast cancer (TNBC) models using Zr-labeled Trodelvy ([Zr]Zr-DFO-Trodelvy).

View Article and Find Full Text PDF

Purpose: The positron range effect can impair PET image quality of Gallium-68 (Ga). A positron range correction (PRC) can be applied to reduce this effect. In this study, the effect of a tissue-independent PRC for Ga was investigated on patient data.

View Article and Find Full Text PDF

Development of a novel molecular probe for visualizing mesothelin on the tumor via positron emission tomography.

Eur J Nucl Med Mol Imaging

January 2025

Institute of Radiation Medicine, Fudan University, Xietu Road 2094, Shanghai, 200032, China.

Objectives: Mesothelin (MSLN) is an antigen that is overexpressed in various cancers, and its interaction with tumor-associated cancer antigen 125 plays a multifaceted role in tumor metastasis. The serum MSLN expression level can be detected using enzyme-linked immunosorbent assay; however, non-invasive visualization of its expression at the tumor site is currently lacking. Therefore, the aim of this study was to develop a molecular probe for imaging MSLN expression through positron emission tomography (PET).

View Article and Find Full Text PDF

Diagnostic Feasibility of Cardiac PET CT in the Evaluation of Inflammatory, Infectious, and Malignant Heart Disease.

R I Med J (2013)

February 2025

Alpert Medical School of Brown University, Department of Medicine, Division of Cardiology, Rhode Island Hospital.

Cardiac Positron Emission Tomography (PET) is a power- ful imaging tool with diverse applications in the detection and diagnosis of various cardiac conditions, including inflammatory, infectious, and neoplastic processes. Using the radiotracer 18F-fluorodeoxyglucose (18F-FDG), cardiac PET enables the identification of cardiac involvement in diseases such as sarcoidosis and severe infections affecting the heart tissue. Additionally, 18F-FDG PET is valuable in the evaluation of cardiac masses, helping to assess their metabolic activity and potential malignancy.

View Article and Find Full Text PDF

Cardiac Positron Emission Tomography (PET) can be used for the assessment of myocardial perfusion. Compared to other cardiac imaging techniques, notably Single Photon Emission Computer Tomography (SPECT), cardiac PET offers superior image resolution, higher accuracy, quantitative measures of myocardial perfusion, lower radiation exposure, and shorter image acquisition time. However, PET tends to be costlier and less widely available than SPECT due to the specialized equipment needed for generating the necessary radiotracers.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!