CT colonography: single- versus multi-detector row imaging.

Radiology

Department of Diagnostic Radiology, Mallinckrodt Institute of Radiology, 510 S Kingshighway Blvd, St Louis, MO 63110, USA.

Published: May 2001

Purpose: To compare respiratory artifacts, colonic distention, and polyp detection at computed tomographic (CT) colonography by using single- and multi-detector row helical CT systems.

Materials And Methods: A total of 237 consecutive patients received subcutaneously administered glucagon and underwent prone and supine CT colonography with single-detector row CT (n = 77) and multi-detector row CT (n = 160), followed by colonoscopy. Examination results were graded for colonic distention, respiratory artifacts, and polyp depiction by two radiologists working independently.

Results: Suboptimal colonic distention was significantly more common with single-detector row CT and was present in at least one segment in 52% (40 of 77 patients) of examinations versus only 19% (30 of 160 patients) with multi-detector row CT (P <.001). Mild respiratory artifacts were present in 61% (47 of 77 patients) of single-detector row CT examinations versus only 16% (26 of 160 patients) of multi-detector row CT examinations (P <.001). Depiction of polyps larger than 10 mm was 89% (eight of nine polyps) for single-detector row CT and 80% (eight of 10 polyps) for multi-detector row CT (P >.05).

Conclusion: CT colonography performed with multi-detector row CT significantly improved the demonstration of colonic distention and depicted fewer respiratory artifacts compared with single-detector row CT. No significant differences in the depiction of polyps larger than 10 mm were demonstrated between single- and multi-detector row CT for a small number of polyps. Studies with a larger prevalence of clinically important polyps are needed for further evaluation of differences in polyp detection.

Download full-text PDF

Source
http://dx.doi.org/10.1148/radiology.219.2.r01ma28461DOI Listing

Publication Analysis

Top Keywords

multi-detector row
16
colonic distention
12
colonography single-
8
respiratory artifacts
8
single-detector row
8
row
6
single- versus
4
multi-detector
4
versus multi-detector
4
row imaging
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!