Atrial fibrillation is often initiated by atrial premature beats originating in the pulmonary veins. Non-invasive localization of these ectopic beats would be of significant value in guiding therapy. Body surface potential mapping was performed in nine patients undergoing invasive electrophysiologic study. Signals were recorded from 62 electrodes during pace mapping from each of the pulmonary veins. Optimal electrodes for localizing pulmonary vein activation were sequentially chosen. Seven optimal electrodes (6 anterior, 1 posterior) for recording ectopic atrial activation originating in the pulmonary veins were selected. The seven optimal electrode set performed better than the standard 9 electrode ECG at estimating the full body surface map (correlation 97 vs. 95.7%; p < 0.05). Seven optimally selected electrodes can estimate the body surface potential distribution during ectopic atrial activation orignating from the pulmonary veins. The ability of this electrode configuration to discriminate the site of origin of ectopic atrial beats requires prospective evaluation.